32
MBW 473 Manual English V5.0
Mirror Cleaning
Activating the Mirror Cleaning function with the respective key at the bottom of the screen will heat the
mirror to a pre-specified temperature, getting the measuring head ready for the removal of the cover
and the optical module. If the mirror and other internal measuring head components are disassembled
while they are cold and become exposed to normal atmospheric air, the possibility of undesired con-
densation exists. Warming the mirror and other internal components to a safe head removal tempera-
ture, greater than or equal to the current ambient temperature, will prevent the formation of dew on the
mirror assembly during servicing.
To set the Minimum Mirror Temperature when activating the Mirror Cleaning mode:
1. Select the
Control Setup
menu by
pressing the dark gray menu selection
key until
Control Setup
appears.
2. Touch the
Mirror Cleaning
menu button.
3. Touch the
Change To:
field to the right
of the
Minimum Mirror Temp
label.
4. Enter the temperature in degrees C which the mirror must warm to during the Mirror Cleaning
mode. It is recommended that you enter your current ambient temperature or higher.
5. If you make an entry error, touch the field that holds the number you wish to change. Each time
you touch the field, the last digit entered will be deleted.
6. Press the OK button in the dialog box or press Enter on the keypad to confirm the new value.
Press Cancel to leave it unchanged.
Mirror Check
Mirror Check is the process of warming the mirror to evaporate all condensation, looking for the pres-
ence of contamination and accounting for it if necessary, then initiating a new dew or frost point meas-
urement. Mirror Check may be started manually with the fixed
Mirror Check
key, or if enabled, it will
start automatically at pre-specified time intervals.
During a mirror check, whether triggered automatically or manually, the indicator on the fixed
Mirror
Check
key has the following meanings:
Red:
The mirror is heating.
Brown:
The mirror is holding at the programmed Mirror Check Temperature.
Yellow:
The mirror is cooling to re-form the dew or frost layer.
Gray:
The mirror check function is not currently active.
Once the system has re-established a dew or frost layer and become stable, the mirror check function
is completed and the color indicator turns gray.
After the Mirror Check is completed the bar of the Mirror Residue Indicator shows the amount of con-
tamination on the mirror. If the bar covers more than a quarter of the space, we recommend that you
clean the mirror. For instructions on mirror cleaning, please refer to section 8.2
’Mirror Cleaning’.
Содержание 473-RP2
Страница 1: ...Operation and Maintenance Manual V5 0 Dew Point Mirror 473...
Страница 2: ......
Страница 4: ...ii MBW 473 Manual English V5 0...
Страница 8: ...2 MBW 473 Manual English V5 0...
Страница 10: ...4 MBW 473 Manual English V5 0...
Страница 14: ...8 MBW 473 Manual English V5 0 SH2 SHX Fan Fan...
Страница 30: ...24 MBW 473 Manual English V5 0...
Страница 68: ...62 MBW 473 Manual English V5 0...
Страница 69: ...MBW 473 Manual English V5 0 63 11 Drawings 11 1 Display Unit Side View...
Страница 70: ...64 MBW 473 Manual English V5 0 Front View Back View...
Страница 71: ...MBW 473 Manual English V5 0 65 11 2 Measuring Heads RP2 SH2...
Страница 72: ...66 MBW 473 Manual English V5 0 SHX SH3...
Страница 73: ...MBW 473 Manual English V5 0 67...
Страница 74: ...68 MBW 473 Manual English V5 0...