– balanSys REV – Extension Gap First: 3in1
Appendix
7 – Assembly of the Tibial Reference System
Fig. 132
Fig. 133
Assembly of the extramedullary Tibial Reference
System:
1. Move the Ankle Holder (71.02.3035 / 1) onto
the Tibial Distal Connector (71.02.3036 / 2).
2. Move the Aiming Device Distal (71.02.3034 / 3)
onto the Tibial Distal Connector (71.02.3036 / 2).
3. Fix the Ankle Holder (71.02.3035 / 1) with the
Tibial Locking Bolt (77.02.0043 / 5).
4. Leave the screw (4) slightly open.
5. Join the Aiming Device Distal (71.02.3034 / 3)
with the Aiming Device Proximal (71.02.3032 / 6).
6. Affix it with the connecting screw
(77.02.0041 / 7).
7. Insert the Tibial Cutting Guide (79.02.0290 / 8)
onto the Aiming Device Proximal (71.02.3032 / 6).
8. Set the REV Tibial Cutting Guide (79.02.0290 / 8)
to zero and tighten the screw (9) with the
Hexagonal Screwdriver (314.270).
Assembly of the intramedullary Tibial Reference
System:
9. Connect the extramedullary Tibial Reference
System with the Intramedullary Shackle
(71.02.3041 / 10).
10. Fix it with the Screw (77.02.0019 / 11).
11. Slide the assembled Tibia Reference System
over the shaft of the chosen Reamer (79.02.0310
to 79.02.0325 / 12).
12. Align the Tibial Reference System distally to
the second metatarsophalangeal bone.
13. Impact on the bracket (13).
14. The adjustment of the cut level is made with the
aid of the T-screw (14). Use the Reference Plate
(77.02.0031) to check the correct level.