Use the Memory Board (79.02.0637) with the Plug-In
Pegs (79.02.0638) to record the diameter of the last
reamer (in this example: diameter 16 mm, length
140 mm).
For stabilization of the Reamer, slide the Guide Sleeve
(79.02.0510 to 79.02.0525) which fits best over the
shaft until it reaches the level of the tibial osteotomy
(only required for the 140 and 200 mm reamers).
Use the Positioning Fork (79.02.0029) to hold down
the Guide Sleeve.
The Tibial Reference System will be guided by the in-
tramedullary Reamer (79.02.0310 to 79.02.0325).
Remark
The assembly of the Tibial Reference System is ex-
plained in Appendix 7.
Slide the assembled Tibial Reference System over the
shaft of the last used Reamer (79.02.0310 to
79.02.0325).
Fig. 2
Fig. 3
Fig. 4