balanSys REV – Extension Gap First: 3in1 –
15
Fig. 13
Fig. 14
Fig. 15
Assemble the Tibial Template (79.02.0291 to
79.02.0295) with the necessary Trial Tibia Augmen-
tation (79.02.0160 to 79.02.0187). This is achieved
with a click-on mechanism.
In this example (left knee), a 5 mm tibial augmenta-
tion for the lateral compartment has been chosen.
Click on the Offset Graduated Collar (79.02.0258)
onto the Tibial Template (79.02.0291 to 79.02.0295).
Remark
Scale can be read by the operator, 12 o’clock = poste-
rior.
Insert the last used Reamer and check the depths. The
marking must be at least equal with the lower side if
the Tibial Template. If the Reamer is not inserted deep
enough, it is necessary to re-ream.
Put the entire configuration (Tibial Template with the
Offset Graduated Collar and Tibial Augmentation)
with the aid of the Holder Tibial Template (71.34.0196)
over the reamer shaft onto the tibial osteotomy.
Surgical technique
Tibial offset determination and preparation