52
– balanSys REV – Extension Gap First: 3in1
Fig. 114
Fig. 115
Fig. 116
In order to screw the dorsal / posterior augmentation
onto the femur implant, turn the Femur Assembly
Block.
Place the augmentation screws into the correct thread
and tighten it with the Allen Key (314.140).
Switch to the Assembly Device. Insert the Reference
Plate (77.02.0031) in the slot of the black Assembly
Device and insert the distal end of the chosen balan-
Sys REV Stem into it. Tighten the fixation screw of the
black Assembly Device.
Remark
There is a marking on the stem – this marking must be
orientated towards the Reference Plate (see picture).
Put the balanSys REV Femur onto the tapered cone of
the stem.
Remark
The taper needs to be dry and free of any foreign
matter before assembling.