balanSys REV – Extension Gap First: 3in1 –
53
Fig. 117
Fig. 118
Fig. 119
Assemble the Offset Alignment Disk (79.02.0287) and
the Box Adapter (79.02.0288). Fit the Box Adapter
into the box of the femur component.
The pictogram on the upper surface highlights the
exact position. The 12 o’clock position must face to-
wards anterior.
Use the Offset Alignment Disk to set the measured
rotation.
Insert the stem screw onto the Screw Positioner
(79.02.0270) and insert the screw into the hole, but
do not tighten it yet!
Remark
One stem screw is packed together with each implant.
The screw is partially turned into the thread of the
stem.
Rotate the Offset Alignment Disk until the Reference
Plate corresponds with the predetermined rotation.