balanSys REV – Extension Gap First: 3in1 –
39
Fig. 80
Fig. 79
Fig. 78
Fig. 77
Press the corresponding Femur Box Inlay (79.02.0470
to 79.02.0474) into the opening of the Trial Femur.
The cam of the box should face towards posterior.
The surface of the Trial Femur and the Femur Box Inlay
must align smoothly.
Assemble the Offset Alignment Disk (79.02.0287) and
the Box Adapter (79.02.0288). The pictogram on the
lower surface of the disk highlights the exact position
of the adapter.
Fit the Box Adapter into the box of the Trial Femur.