Generation II Entree Bath

  All Models 

OPERATION MANUAL 

  

07-01-2018 

Page 27 of 50

 

2071 14

th

 Ave – Columbus, NE 68601 

800-798-5867  FAX: 402-563-9102 

www.mastercarebath.com

 

Cleaning Procedures: 

Turn on BathAire

(SM)

.

 

Push down on the Door Close button until 

the door is fully closed  and stops on its own. Make sure the 

drain is open and  use the shower wand  to rinse all inside 

surfaces and the seal area of the tub to remove residual organic 

materials left over from the previous bath. Close the drain. Turn 

on the water and fill the footwell with approximately one (1) 

gallon of water.  Using  MasterCare's  specially formulated 

Disinfectant  cleaner, dispense 2  oz  (approximately  16 pumps) 

into the foot well water. Using the special Long  Handled  Tub 

Brush, scrub away any remaining organic materials sticking to 

the inside tub surfaces, comfort pads  and belt. Scrub the seal 

area with the Seal Brush  also  included in the tub kit.  Leave 

disinfectant on surface for a minimum of 10 minutes per 

manufactures requirements. Open the drain. Rinse all the inside 

surfaces and seal area with clean water from the shower wand 

at full pressure. Turn off BathAire

(SM

)

.

 

  Open the door all the 

way. Allow all cleaned surfaces to dry naturally with the drain 

open. The clean comfort pads can be placed in to the bottom of 

the foot well to dry. 

A) Never leave individual unattended during the entire 

bathing procedure. B) All individuals must be assessed 

before using the system, to determine if they will require 

the assistance of two staff members to transfer into and 

out of the bath or if one staff member is suitable for this procedure. 

Individuals identified as needing the assistance of two staff members 

must always be transferred with two staff members. 

BathAire(SM) Purge Cycle: 

BathAire

(SM)

 will automatically turn on after the bath has been 

idle for twenty (20) minutes and run for one (1) minute. This 

will help the area around the aerators to dry. 

Other Considerations:  
CAUTION

When not being used,  all Disinfectant Cleaner, 

Bath Oil and Shampoo solutions used during the bathing 

procedures must be stored in a secure area not accessible by 

the patients. 

It is recommended that a two person bathing team be 

established for each shift with trained backups. Using the 

team approach, with one member transferring the individuals 

to and from the bathing area and the other operating the 

bathing system,  increases the efficiency of the bathing 

operations. 

It is recommended that the operations wall chart be posted 

on the wall next to the control end of the tub for quick 

reference by the operator of the Entree Bath system (One is 

provided in the tub starter kit). 

MasterCare 

BathAire System 

Disinfectant Cleaner 

Item: 90101 

Long Handled Tub Brush 

BathAire 

(SM)

 Aerator 

CAUTION:

 

Failure to leave the door OPEN on 

the tub when not in use will cause damage 

to the seal and void the warranty for this 

item.

  

Seal Brush 

Содержание Gen II Entree Bath

Страница 1: ...All Instructions Thoroughly Important safety instructions Read and follow all instructions thoroughly CAUTION The caution heading identifies important maintenance and operation information Failure to...

Страница 2: ...eneration II Entree Bath All Models OPERATION MANUAL 07 01 2018 Page 20 of 50 2071 14th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com This page is intentionally left blank...

Страница 3: ...ir dryer lamp telephone radio or television within 5 ft 1 5m of the hydromassage bathtub Proper electrical grounding must be performed by a licensed electrician following local state and federal codes...

Страница 4: ...ll and Shower Spray Control Valves Turns ON OFF and controls the flow of mixed water to the tub Fill Spigot and Shower Wand Solar Powered Thermometer Digital readout temperature of the mixed water to...

Страница 5: ...utton once and BathAire SM will turn on Press the button once again and the BathAire SM will shut off BathAire SM More Less Air Heat Buttons Once the BathAire SM is on it is in an adjustment mode Bath...

Страница 6: ...out of the way Position the comfort pad group as follows two rectangular pads parallel about three to four inches apart centered in the individual seating area and one rectangular pad horizontally on...

Страница 7: ...the Dispensing bottle This will help reduce foaming from soap shampoo etc CAUTION If there is water in the tub the drain sensors will not let the Door Switch be activated This will prevent accidental...

Страница 8: ...BathAire SM system 10 Dry as much of the individual as you can before removing them from the tub by leaning the individual forward and towel dry their back side and inside of the tub area behind the...

Страница 9: ...ttended during the entire bathing procedure B All individuals must be assessed before using the system to determine if they will require the assistance of two staff members to transfer into and out of...

Страница 10: ...fer device is needed 11 Push down on the Door Open Button allowing the door to rotate to a fully open position until it stops on its own 12 Remove the Kick Panel by removing the two thumb screws at ei...

Страница 11: ...cases the door can be opened with the yellow handled 5 16 ball end Allen wrench driver see also GFCI Reset Procedures 3 Remove the emergency door open close access panel by removing the two thumb scr...

Страница 12: ...and reset using the following procedures GFCI Ground Fault Circuit Interrupter detects and interrupts electric faults to insure safety to all individuals Reset Button trips and shuts down all power sh...

Страница 13: ...e pad as far as they go Pull gently up on one side until all the cups release CAUTION Do not pull up on pad by grasping it on the outside and trying to pull up on all suction cups at the same time wit...

Страница 14: ...eneration II Entree Bath All Models OPERATION MANUAL 07 01 2018 Page 32 of 50 2071 14th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com This page is intentionally left blank...

Страница 15: ...Adjust Mixing Valve To Desired Temperature Fill Tub with Tub Fill Valve Use Liquids As Desired Use Shower Wand With Shower Wand Control Valve Open Drain With Open Close Lever After Water Is Drained O...

Страница 16: ...eneration II Entree Bath All Models OPERATION MANUAL 07 01 2018 Page 34 of 50 2071 14th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com This page is intentionally left blank...

Отзывы: