EMERGENCY OPEN/CLOSE DOOR PROCEDURES 

1.

 

In the event of a medical emergency

 where it is needed to open the 

door before the tub is totally empty: hold Emergency Door Open 

Button and the Door Open Button 

at the same time 

to open the 

door

It will override the drain water sensors. 

2.

 

In the event there is a power failure

 or a failure in the door actuator 

system, first check to make sure the drain does not have any water. If 

there is water in the drain, the drain sensors will deactivate the Door 

Open Button. In most cases, the door can be opened with the yellow 

handled 5/16" ball end Allen wrench driver. 

(see also GFCI Reset 

Procedures.) 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3.

 

Remove the emergency door open/close access panel by removing the two thumb screws at either end of the panel. 

Set the panel and thumb screws off to the side. 

4.

 

Locate the yellow handled 5/16” ball end allen wrench nested in a hole under the access panel. 

5.

 

Engage the ball end allen wrench through the open hole in the tub deck into the socket at the top of the actuator 

screw shaft. 

6.

 

Turn the handle 

clockwise

 to lower the door. 

7.

 

Replace the Allen wrench and access panel when the power is restored.

Emergency Open  

T-Handle Allen Wrench 

Access Panel 

T-Handle 

Allen Wrench 

Ball-End of 5/16" 

Allen Wrench 

Socket Head for 

5/16" Allen Wrench 

Top of Actuator 

Screw Shaft 

Thumb Screw 

Thumb Screw 

FOOT END CONTROL PANEL 

Door Open 

button

 

Emergency 

Door Open 

button

 

Содержание Gen II Entree Bath

Страница 1: ...All Instructions Thoroughly Important safety instructions Read and follow all instructions thoroughly CAUTION The caution heading identifies important maintenance and operation information Failure to...

Страница 2: ...eneration II Entree Bath All Models OPERATION MANUAL 07 01 2018 Page 20 of 50 2071 14th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com This page is intentionally left blank...

Страница 3: ...ir dryer lamp telephone radio or television within 5 ft 1 5m of the hydromassage bathtub Proper electrical grounding must be performed by a licensed electrician following local state and federal codes...

Страница 4: ...ll and Shower Spray Control Valves Turns ON OFF and controls the flow of mixed water to the tub Fill Spigot and Shower Wand Solar Powered Thermometer Digital readout temperature of the mixed water to...

Страница 5: ...utton once and BathAire SM will turn on Press the button once again and the BathAire SM will shut off BathAire SM More Less Air Heat Buttons Once the BathAire SM is on it is in an adjustment mode Bath...

Страница 6: ...out of the way Position the comfort pad group as follows two rectangular pads parallel about three to four inches apart centered in the individual seating area and one rectangular pad horizontally on...

Страница 7: ...the Dispensing bottle This will help reduce foaming from soap shampoo etc CAUTION If there is water in the tub the drain sensors will not let the Door Switch be activated This will prevent accidental...

Страница 8: ...BathAire SM system 10 Dry as much of the individual as you can before removing them from the tub by leaning the individual forward and towel dry their back side and inside of the tub area behind the...

Страница 9: ...ttended during the entire bathing procedure B All individuals must be assessed before using the system to determine if they will require the assistance of two staff members to transfer into and out of...

Страница 10: ...fer device is needed 11 Push down on the Door Open Button allowing the door to rotate to a fully open position until it stops on its own 12 Remove the Kick Panel by removing the two thumb screws at ei...

Страница 11: ...cases the door can be opened with the yellow handled 5 16 ball end Allen wrench driver see also GFCI Reset Procedures 3 Remove the emergency door open close access panel by removing the two thumb scr...

Страница 12: ...and reset using the following procedures GFCI Ground Fault Circuit Interrupter detects and interrupts electric faults to insure safety to all individuals Reset Button trips and shuts down all power sh...

Страница 13: ...e pad as far as they go Pull gently up on one side until all the cups release CAUTION Do not pull up on pad by grasping it on the outside and trying to pull up on all suction cups at the same time wit...

Страница 14: ...eneration II Entree Bath All Models OPERATION MANUAL 07 01 2018 Page 32 of 50 2071 14th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com This page is intentionally left blank...

Страница 15: ...Adjust Mixing Valve To Desired Temperature Fill Tub with Tub Fill Valve Use Liquids As Desired Use Shower Wand With Shower Wand Control Valve Open Drain With Open Close Lever After Water Is Drained O...

Страница 16: ...eneration II Entree Bath All Models OPERATION MANUAL 07 01 2018 Page 34 of 50 2071 14th Ave Columbus NE 68601 800 798 5867 FAX 402 563 9102 www mastercarebath com This page is intentionally left blank...

Отзывы: