Test voltage:
4V DC (into open circuit)
Test current:
200mA DC (into short circuit)
Accuracy (referenced to 230V rms):
± (5% r 2 digits)
Threshold for ‘Simulated Current Hi’:
16A
Memory
Completed test types that can be stored:
200
General
Overload protection:
300V AC/DC
Protection class:
IP20
Safety:
Complies to BS EN 61010-1
Temperature Range:
-10 to 40°C, non-condensing
Batteries:
6 x 1.2V AA NiMH rechargeable batteries
1800mAh, or greater
Caution - use only the battery type specified
Battery Life:
Fully charged, minimum 600 complete tests, stored
to memory. The HPAT600 will operate fully from the battery
charger in the event of discharged batteries. Do not remove
batteries.
Auto Power Down:
After 10 minutes of inactivity, or when battery low
Battery Charger:
Input:
230V 50Hz
Output:
12VDC at 300mA, 5.5 jack x 2.5 bore centre
positive
Dimensions:
90 x 210 x 54mm
Weight:
Approx 700g (including batteries & leads)
Specification
HPAT600
HandyPAT
Содержание HPAT600/2 HANDYPAT
Страница 1: ...HPAT600 2 HANDYPAT Instruction Manual ...