u
Connect the test probe to the HPAT600 4mm test
probe socket (2) and connect the other end to any
exposed metal work on the appliance.
u
If the appliance being tested has more than one
instance of exposed metalwork, it is advisable to
connect the test probe to each in turn and perform
additional individual Class II insulation tests. Refer to
3.9 (Class II Insulation Test)
u
Where the appliance being tested has one or more
areas of possible insulation breakdown, such as
cooling slots or cover joints, it is advisable to wrap
metal foil tightly around the appliance and connect
the test probe to the metal foil then perform Class II
insulation tests.
Class II Test
Connection
Testing in Manual Mode
u
Press
CLASS II
to select the Class II appliance test
and display the diagram screen.
u
Press
OK
for the HPAT600 to display the VISUAL
CHECK screen.
u
If any visual or fuse checks fail, select
Fail
and press
OK
to abort the test and return to the select test
screen.
23