Result of Test
u
At the end of the test the HPAT600 will display the
value of the simulated load test current and the
calculated power. These values are referenced to a
simulated mains supply voltage of 230V and a unity
power factor.
u
If the HPAT600 displays the
‘Simulated load current
Hi’
message, refer to 3.13. (Simulated Load Current
Hi)
u
Test data can now be saved (
SAVE
) to memory if
required. Refer to 2.14. (Saving Test Data to Memory)
3.7 Class I Earth Continuity Test
Always refer to and perform 3.2 before proceeding
This is a single test to measure resistance between earth
and exposed metalwork.
Test Connections
u
Refer to the Class I Earth Continuity connection
diagram below.
u
Plug the appliance mains plug into the HPAT600 front
panel mains socket (1).
u
Connect the test probe to the HPAT600 4mm test
probe socket (2) and connect the other end to any
exposed metal work on the appliance.
u
If the appliance being tested has more than one
instance of exposed metalwork, it is advisable to
connect the test probe to each in turn and perform an
earth continuity measurement.
26