This data can now be saved (SAVE) to memory if
required. Refer to 2.14. (Saving Test Data to Memory).
Note: If the earth continuity or insulation measurements
FAIL, then Repeat and Abort are the only options.
For example if a Class I test fails earth continuity the
HPAT600 will not proceed to the insulation measurement.
3.13 Simulated Load Current Hi
u
If the simulated load test indicates a load current
value > 16A, the ‘Simulated load current Hi’ message
is displayed.
u
If the simulated load test current value is > 30A, the
display will flash ‘30.0’.
u
The HPAT600 is designed for the testing of appliances
that are powered from 13A mains sockets, so if the
‘Simulated load current Hi’ message is displayed it
would be advisable to have the appliance investigated
by a suitably qualified technician.
3.14 Testing of 110V appliances
A TL157 PAT 110V Adaptor (available separately) is
required for testing 110V appliances (yellow BS4343
16A 110V plug).
Refer to the connection diagram below.
TL157 110V
Adaptor