Result of Test
u
At the end of the test the HPAT600 will display
‘PASS’ or ‘FAIL’ to indicate the result of the test. The
measurement result is also displayed.
u
Test data can now be saved (SAVE) to memory
if required. Refer to 2.14. (Saving Test Data to
Memory)
3.10 Extension Lead Testing
Always refer to and perform 3.2 before
proceeding
To test extension leads the HPAT600 IEC Lead Test is
used with connections as detailed below. Refer to 3.3
(IEC Lead Test).
A Martindale EX332 IEC mains lead (or similar) is
required to perform extension lead tests.
Test Connections
u
Refer to the Extension Lead connection diagram
below.
u
Plug the extension lead plug into the HPAT600 front
panel mains socket (1).
u
Plug the EX332 IEC lead mains plug into the
extension lead socket.
u
Connect the other end of the EX332 IEC lead to the
HPAT600 IEC input socket (3).