Extension Lead
Connection
3.11 Fuse Check/Simulated Load Failure
Possible Reasons for Failure other than a Faulty
Appliance or Ruptured Fuse
u
Appliance mains switch not set to the ON position.
u
Appliance not correctly connected to the HPAT600.
u
Some types of appliance, for example those with
switched mode power supplies, will indicate ‘FAIL’
for the fuse check and simulated load test due to the
nature of their circuitry.
A ‘FAIL’ in such circumstances does not necessarily
mean a fuse is ruptured or the appliance is faulty.
Remove the fuse/fuses and check fuse continuity by an
alternate means.
Ensure the appliance is NOT plugged into the mains
supply when removing fuses.
Continue Test
If it is determined the fuses are satisfactory and the
appliance was of a type that would fail the fuse check,
Class I or Class II appliance tests can be continued by
selecting
Continue
and pressing
OK
to proceed to the
next step of the test.
32