3. Maintenance
The operating table will have to be serviced once
a year by the MAQUET factory service department
or by service technicians authorized by MAQUET
to do so.
Please adhere to prescribed maintenance intervals as
otherwise guarantee coverage will lapse.
In so doing you also make a contribution to patient and
staff safety and extend the useful life of your costly equip-
ment.
We recommend concluding a maintenance contract with
MAQUET.
Outside Germany please contact your local representa-
tive in all matters relating to service.
4. Malfunctions
If the operating table is defective, we recommend pro-
ceeding as follows in order to place the operating table
on its castors and remove it from the operating room:
1. Use the UNLOCK function on the Override panel: press
the ON key and the UNLOCK key simultaneously
until the table is standing on its castors.
If required:
2. Use the mains cable to connect the operating table
to the mains supply and repeat step 1.
If required:
3. Use the ratchet:
•
Remove the sealing cap at the base using a wide
bladed screw driver or a coin.
•
Insert the ratchet (ref. no. 3113.3649, included in
the delivery) into the opening.
•
Turn the ratchet (2) clockwise until the castors
are extended (visual check). When the castors
are fully extended, it will be very difficult or im-
possible to turn the ratchet (2).
•
Screw in the sealing cap without tightening.
•
Move the defective operating table to the desired
location and park it on its castors (secure it, if
required)
•
Inform the service department.
Attention:
•
If the castors have been extended using the hand
crank, the UNLOCK key must not be actuated, oth-
erwise the base will be lowered and cannot be placed
on its castors again.
If the operating table malfunctions, please contact the
MAQUET agency. If you can describe the symptoms
and have the system’s serial number ready, they will be
able to react and solve the problem more quickly.
Î
Î
Î
Î
Î
Note
In case of malfunctions, do not attempt to repair the
unit yourself and do not try to force the unit.