25
1133.02BX
GA113302GB11
2. Positioning a patient weighing up to 135 kg (300 lb)
2.1 Approved accessories
Danger of injury!
The mounted accessories may not exceed the maximum permissible length specified in the sketch.
For further accessories, please refer to the accessories list, see Chapter VIII.
2.2 Operating table unlocked (UNLOCK)
Danger of tipping the operating table over!
If the patient is in the wrong position, the operating
table may tip over when the castors are deployed
(UNLOCK).
Position the patient like Fig. V-1 shows and set
the operating table to a height of less than
940 mm before deploying the operating table
castors (UNLOCK). The position below 940 mm
is reached when the warning sign on the column
is no longer seen.
V. Patient positioning
NO
YES