MODEL YB8
MAINTENANCE
April 1999/Manning Environmental Inc.
Page C-7
EPA-600/4-77-039 ("Sampling of Water and Wastewater" by Dr. Phillip E. Shelley), or consult with the
facility that will do the actual testing of the samples. They could probably assist in setting a cleaning regime
that will help produce the most accurate results possible.
The following procedures are very general outlines of procedures for cleaning certain parts of the sampler:
Intake Hose
There are two types of intake hose used with the sampler - PVC and
Teflon . PVC intake hose is used for general purpose sampling
®
(Non-Toxic) applications. Teflon hose is used for priority pollutant
®
sampling (Toxic) applications.
1.
Remove the intake hose. Remove the strainer if
necessary.
2.
Wash the intake hose and strainer using a
cleaning solution appropriate for the application.
The use of methylene chloride or other solvents
may leave a residue that could contaminate the sample. Use a test tube brush to scrub the
internal surfaces of the strainer. Pull the brush through the hose with a wire to clean the
internal surfaces of the hose.
3.
Rinse the hose and strainer thoroughly in clean water (warm water is best) and
reassemble.
It may be easier and more convenient to simply use a new hose for each sample configuration. This
eliminates cleaning and disposal of potentially hazardous regulated chemicals.
Liquid Sensor
The continuity type liquid sensor will need to be cleaned occasionally. Like any piece of the sample tract, it
has the ability to collect particulate matter, which needs to be
removed. By removing this particulate matter, it ensures that you will
receive optimum performance from the unit.
The ultrasonic liquid sensor will not need to be cleaned. By changing
the pump tubing you in effect, clean, the ultrasonic sensor.
1.
Remove the pump tubing from the liquid
connectors.
2.
Insert a bottle brush or other scrubbing type
device and vigorously insert the scrubber in and out of the sensor.
3.
Reconnect the pump tubing to the liquid connectors and run a manual cycle with clean
Содержание YB8
Страница 2: ......
Страница 4: ......
Страница 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Страница 30: ......
Страница 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Страница 90: ......
Страница 106: ......
Страница 108: ......
Страница 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...