![Manning YB8 Скачать руководство пользователя страница 65](http://html1.mh-extra.com/html/manning/yb8/yb8_operating-and-instruction-manual_701729065.webp)
MODEL YB8
PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
*02 operates much like basic FLOW mode except a time override is added. The override time ensures a
sample is collected, after a set amount of time has elapsed, if a contact closure has not been received. Once
the program has been started the sampler will immediately begin counting down the override time. The
sampler is also concurrently waiting for contact closures. Every time a contact closure is detected the sampler
will advance the spout, draw a sample, and then place it in a bottle. It will then wait for the next contact
closure while displaying a running tally indicating the number of samples collected to that point. If there have
been no contact closures by the end of the override time the user specifies, the sampler will advance the spout,
draw a sample, and then place it in a bottle. The override timer will then reset and immediately start counting
down again while waiting for the next contact closure. This will continue until the total number of bottles the
unit is configured for (set in *99) each have their set number of samples placed in them or a bottle full
condition occurs. If either of these two conditions occur, the sampler ends the program.
he unit will take 1 sample per override time interval until all bottles have samples.
Display on LCD
Explanation
SAMPLER READY
04/30
04:30:02
This display indicates the sampler is ready to program and displays
the current time. Press the
*
key to begin programming.
ENTER * MODE
__ __
04:30:02
The sampler is now prompting for a star mode to be input. Enter the
numbers which represent the star mode of choice, in this example 02,
and press <ENTER>.
TIME OVERRIDE?
__ __ : __
__
04:30:02
Input the maximum time the sampler will be allowed to wait to
receive a contact closure, understanding that if the unit has not
received a closure in this time, it will automatically take a sample.
Press <ENTER> after inputing.
PUSH START/OPTIONS
04:30:02
If no add-on options are desired, press <START> to begin the
program.
FLOW MODE (* 02)
__ __ __ __
04:30:02
The sampler is now ready to receive contact closures and is
independently counting down the Time Override.
*02 Flow Mode - Time Interval Override
Содержание YB8
Страница 2: ......
Страница 4: ......
Страница 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Страница 30: ......
Страница 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Страница 90: ......
Страница 106: ......
Страница 108: ......
Страница 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...