PROGRAMMING SECTION
MODEL YB8
April 1999/Manning Environmental Inc.
3DJH%
Add-On Programming Functions
Multiple Bottles per Sampling Event
Multiple Bottles per Sampling Event is not a stand alone function but works in conjunction with TIME,
FLOW and certain * Modes (*02, *04, *05, *06, *09) to expand the capabilities of the sampler. This option
places 1 sample in from 2-24 bottles in rapid succession during 1 sampling event (such as a contact closure or
a time interval). To use Multiple Bottle per Sampling Event, the user selects the mode of choice - a TIME,
FLOW, or * Mode. After entering the required information, the PUSH START/OPTIONS prompt will
appear on the display. At this cue press <MULTI BOTTLE> and enter the number of bottles into which 1
sample should be placed, in rapid succession. For example, Multiple Bottles per Sampling Event would be
used if a sample is to be taken every 100,000 gallons, and 1 sample is to be placed in 5 different bottles each
time there is a contact closure. The unit would start the sampling sequence after it had received a contact
closure. It would draw and place 1 sample in the first bottle, immediately move to the second, draw and place
1 sample in that bottle, immediately move to the third, draw and place 1 sample in that bottle and so on until it
had deposited 1 sample in the specified number of bottles (in this example 5).
Display on LCD
Explanation
SAMPLER READY
04/30
04:30:02
This display indicates the sampler is ready to program. Multiple
Bottles per Sampling Event is not a stand alone program. It works in
conjunction with FLOW, TIME, and certain * Modes. In this
example <FLOW> was pushed as the mode of choice.
PUSH START/OPTIONS
04:30:02
At this prompt the user selects Multiple Bottles per Sampling Event,
by pressing <MULTI BOTTLE>.
BOTTLES PER SAMPLE?
__ __
04:30:02
Input the number of bottles into which 1 sample will be placed in
rapid succession during a sampling event and press <ENTER>.
PUSH START/OPTIONS
04:30:02
At this point, simply press <START>. NOTE: Multiple Bottles per
Sampling Event and Multiple Samples per Bottle cannot be selected
simultaneously. They are mutually exclusive options.
FLOW MODE
__ __ __ __
04:30:02
The unit is now waiting for a contact closure to initiate the sample
sequence.
Содержание YB8
Страница 2: ......
Страница 4: ......
Страница 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Страница 30: ......
Страница 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Страница 90: ......
Страница 106: ......
Страница 108: ......
Страница 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...