MODEL YB8
PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
Display on LCD
Explanation
SAMPLER READY
04/30
04:30:02
This display indicates the sampler is ready to program and displays
the current time. Press the
*
key to begin programming.
ENTER * MODE
__ __
04:30:02
The sampler is now prompting for a star mode to be input. Enter the
numbers which represent the star mode of choice, in this example 09,
and press <ENTER>.
UPPER LEVEL LIMIT?
__ __ __ __
04:30:02
Enter the Upper Level Limit as a 4-digit number. Remember the
decimal is implied in this program and the unit of measure is generic,
so if the entry were to be 10 feet/meters/inches/millimeters, it could
be entered as 0010, 0100, or 1000. Be CONSISTENT in all entries
throughout the program.
LOWER LEVEL LIMIT?
__ __ __ __
04:30:02
Enter the Lower Level Limit as a 4-digit number. Remember to be
consistent with the implied decimal from previous entries.
RISING DELTA?
__ __ __ __
04:30:02
Enter a 4-digit number which represents the rising change in water
level that will trigger a sample to be taken. If the rise of the water is
equal to or greater than this number a sample will be taken.
FALLING DELTA?
__ __ __ __
04:30:02
Enter a 4-digit number which represents the falling change in water
level that will trigger a sample to be taken. If the fall of the water is
equal to or greater than this number a sample will be taken.
SAMPLING LEVEL 1?
__ __ __ __
04:30:02
Enter a 4-digit number that represents the lowest level at which a
sample is to be taken. Must be greater than the Lower Level Limit.
Remember to be consistent with the implied decimal from previous
entries.
TIME OVERRIDE 1?
__ __ : __ __
04:30:02
Enter a time in HH:MM format. This is the amount of time after
Sampling Level 1 during which the sampler waits for an event
(Rising or Falling Delta, Sampling Level 2 reached, etc.). If no event
occurs before the interval is done, a sample will be taken. If an event
occurs, the Time Override will reset, or move to Time Override 2 if
Sampling Level 2 has been reached.
SAMPLING LEVEL 2?
__ __ __ __
04:30:02
Enter a 4-digit number representing the next level at which a sample
should be taken, keeping consistent with the implied decimal point in
previous entries. A sample will be taken when the water level rises to
this point. Must be greater than Sampling Level 1.
TIME OVERRIDE 2?
__ __ __ __
04:30:02
Enter a time in HH:MM format. Operates on the same principal as
Time Override 1.
Содержание YB8
Страница 2: ......
Страница 4: ......
Страница 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Страница 30: ......
Страница 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Страница 90: ......
Страница 106: ......
Страница 108: ......
Страница 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...