![Manning YB8 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/manning/yb8/yb8_operating-and-instruction-manual_701729047.webp)
MODEL YB8
PROGRAMMING SECTION
April 1999/Manning Environmental Inc.
3DJH%
ENTER SCROLL SECONDS
__ __
04:30:02
To set the scroll seconds, enter a 2 digit number representing the
amount of time, in seconds, you wish the display to show a recorded
event before advancing to the next screen. After inputting press
<ENTER>
The sampler displays recorded events in one of two ways:
A. By entering the scroll seconds, the sampler automatically
advances sequentially through the recorded events, showing each
event for the set number of scroll seconds. This will continue until
the event entered in the COUNT # is displayed. The sampler will
then return to the ENTER SELECTION prompt within *91.
B. The user can also manually review the logged events, although
scroll seconds still have to be entered. To manually examine the
logged events press <DISPLAY> once for each event to be
reviewed. If <DISPLAY> is not pressed, the unit will default and use
the entered scroll time to advance the display.
PUSH THE DISPLAY
KEY FOR NEXT EVENT
This is a momentary display (3 seconds) to remind the user that they
can manually advance the log review or that the unit will do it
automatically based on the time set at the scroll seconds prompt.
_ _:_ _:_ _ , E_ _ _ , B_ _
S_ _ _ , _ _, _ _
This display is divided into multiple sections to communicate
information about the logged sample.
1 Line
st
1 section - Time at which the sample was collected.
st
2 section - Headed by a capital "E", indicates the event number.
nd
3 section - Headed by a capital "B" represents the bottle number.
rd
2 Line
nd
1 section - Headed by a capital "S" indicates the sample number.
st
2 section - Trigger Codes - This is a 2 letter code that specifies
nd
what initiated the sample. For a complete list of codes,
refer to page 11 - TRIGGER CODES.
3 section - The last section signifies result code. This tells the user
rd
whether the sampler was successful or unsuccessful in
collecting a sample and why.
To quit viewing data, simply press <RESET> once. This takes you to
the beginning of the menu selection in the Data Logging menu. The
unit will continue to show the events either based on the scroll time or
by pressing <DISPLAY> until the STOP # is reached. At this point
the unit will return the operator to the ENTER SELECTION prompt.
Содержание YB8
Страница 2: ......
Страница 4: ......
Страница 22: ...INSTALLATION AND OPERATION MODEL YB8 April 1999 Manning Environmental Inc Page A 12...
Страница 30: ......
Страница 61: ...MODEL YB8 PROGRAMMING SECTION April 1999 Manning Environmental Inc THIS PAGE BLANK...
Страница 90: ......
Страница 106: ......
Страница 108: ......
Страница 119: ...9 YB8 Sampler Theory of Operation Figure 8 YB8 Sampler Block Diagram...