Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
36
30/06/2021
5.4.3.3 Immittance Measurements
Show the probe to the patient and then explain the following:
•
An eartip is placed on the tip of the probe and inserted into the ear canal. A seal
must be achieved for the test to progress.
•
Coughing, talking and swallowing will disturb test results.
•
The aim of
Tympanometry
is to test the mobility of the eardrum and the
condition of the middle ear.
o
A small amount of air will flow through the probe to move the eardrum; it produces
a sensation equal to pressing a finger slightly into the ear canal.
o
One or more tones will be heard during the test. No participation is
expected from the patient.
•
The aim of
Acoustic Reflexes
is to test the condition of the Musculus stapedius.
o
One or more louder tones will be heard during the test. No participation is
expected from the patient.
•
The aim of
Reflex Decay
is test the integrity of the CN VIII.
o
One tone is presented above the acoustic reflex threshold measurement for
a minimum period of 10 seconds.
•
The aim of
ETF
is to test the condition of the Eustachian tube.
o
ETF Intact
: three tympanograms are completed while the patient performs
a manueaver between each tympanogram.
o
ETF Perforated
: pressure level is obtained in the ear canal and the patient
swallows to measure change of pressure.