Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
70
30/06/2021
5.6.12 Settings
– Decay – General (MI 36 Version)
Figure 136
Figure 137
Duration:
Defines the length the tone will be presented
to the patient (Figure 136).
Duration
can be configured
in by 5 second increments from 10 s to 30 s.
Level
: Defines the default intensity of the stimulus upon
entering the screen (Figure 137). You can adjust the
level in 5 dB steps.
NOTE:
The deflection of the acoustic reflex is defined
within the
Acoustic Reflex
settings.
5.6.13 Settings
– ETF – Intact (MI 36 Version)
Figure 138
First Test:
Defines the message while the test is in
progression. The end user can select which maneuver
will be displayed first,
Swallow or Valsalva
NOTE:
The selection also determines the color of the
tympanometry graphs:
•
Swallow
represented by
Orange
•
Valsalva
represented by
Green
Figure 139
Pump speed
: Selection of pump speed determines how
precisely and quickly the test will proceed (Figure 139).
NOTE:
A slow speed is more time consuming, but
may give more detailed information.
There are four different
pump speed
Settings:
•
Automatic
(Dynamic from 600 daPa/s for low gradient
and 200 daPa/s for a gradient larger than 5 daPa)
•
Minimum
(50 daPa/s): slow, very precise results
•
Medium
(250 daPa/s): compromise of speed and precision
•
Maximum
(>400 daPa/s): fast, screening
Figure 140
Auto stop
: will automatically stop measurement when
hitting the zero line to lessen the test time without
affecting results (Figure 140).
Figure 141
Start pressure:
the pressure that is first introduced
when performing
ETF
– Intact
measurement.
Stop pressure:
the end pressure of the
ETF
– Intact
measurement (Figure 141).
NOTE:
You can adjust the pressure in 25 daPa steps.