![Maico touchTymp MI 26 Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/maico/touchtymp-mi-26/touchtymp-mi-26_operation-manual_3686551070.webp)
Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
69
30/06/2021
In general,
AGC
is used to hold the level of the tone constant.
Especially in smaller ear canal volumes
AGC
provides an
accurate and safe intensity reflex stimulation. Without
AGC
,
the reflex activator stimuli in these smaller ear canals would
be higher than the referenced calibration value.
5.6.10 Settings
– Acoustic Reflexes – Level
Figure 130
Figure 131
Figure 132
Figure 133
Level
: Defines the test operation on which level change is used
when entering the
Acoustic Reflex
module. Options include:
•
Automatic
: touchTymp starts
Acoustic Reflex
test at
the minimum level and increases in 5 dB steps
automatically until a reflex is registered or the
maximum level is reached (Figure 130).
•
You can adjust the minimum and maximum level for
Ipsi
and
Contra
in 5 dB steps either for the level range
(if
Automatic
is selected) or a single level (if
Fixed
is
selected). Levels can be selected between:
•
Ipsi
: min: 70 dB HL, max: 105 dB HL,
•
Contra
: min : 70 dB HL , max : 120 dB HL (Figure 131).
•
Fixed:
The measurement is performed at one level as
defined in the
Settings
5.6.11 Settings
– Acoustic Reflexes – Stimulus
Figure 134
Figure 135
Ipsi 226 Hz, Ipsi 1000 Hz, Contra 226 Hz, Contra
1000 Hz:
Defines the default frequencies for
Ipsilateral
and
Contralateral
measurements when the
Acoustic
Reflex
screen is entered for testing. Default frequencies
can be modified within the test screen and will return to
the default settings when the screen has been exited
(Figure 134).
NOTE:
Stimuli for MI 26 version are frequencies
500 Hz, 1000 Hz, 2000 Hz, 4000 Hz
. Stimulus options
for MI 36 version also include noise stimuli (
BB
–
Broadband,
HP
– High Pass,
LP
– Low Pass).
When options are greyed out in the settings screen, the
license is not active (Figure 135).