Operation Manual touchTymp MI 26 and MI 36 Version
8511780 Rev. 4
55
30/06/2021
#
Name(s) /
Function(s)
Description
12
Status Bar
The numerical display for the testing operation of the device
(Figure 88).
Figure 88
The information includes (Table 14):
Table 14 Status Bar explanation
#
Status Bar Explanation
1a/
1b
When a stimulus is presented to the patient the
small box on the edge of the bar turns green. The
left is for the test signal and the right box is for the
masking noise.
2
Stimulus symbol (i.e. pure tone, pulse, etc.) for test ear.
3
Hearing level of test signal, displayed in the color
of the ear being tested (Right ear/Red, Left
ear/Blue).
4
Frequency of test signal.
5
Masking level displayed in the color of the non-test ear.
6
Noise symbol for masking
13
Result Display
Results are stored on the device for later printing or
transferring to the PC. The display can be configured to
Graph
,
Table
or
None
within the settings. See section
5.6.16 for more information.
NOTE
:
None
-setting display does not allow storing of results.
Delete point
Only shown in
Edit mode
(see Table 12). Single
measurement results can be deleted.
1a
2
3
4
5
6
1b
Содержание touchTymp MI 26