Operation Manual easyScreen
8508935 Rev. 4
19
04/04/2019
Cleaning and Disinfecting the Case and Cables
Use caution while cleaning.
Before cleaning, remove the easyScreen from the cradle
and unplug the cradle from AC power.
Use a damp cloth to clean the plastic parts of the
easyScreen and cradle.
For disinfection use a disinfectant wipe rather than a spray
product. Make sure that excess liquid from the wipe does not
seep into any sensitive areas such as connectors,
BERAphone
®
speaker and seams where plastic pieces
connect such as the edges around the touch screen.
Follow the instructions on the disinfection product.
Cleaning and Disinfecting the BERAphone
®
The BERAphone
®
features stainless steel electrodes and a silicone ear cushion
designed to make direct contact with the baby’s skin. Assuming intact skin on the baby,
the device is considered non-critical according to the CDC guidelines for minimizing
cross-infection, meaning that the device contacts the patient only externally on intact
skin. Therefore, sterilization is not required. However, the device must be cleaned and
disinfected with a medical or hospital-grade disinfectant prior to re-use.
Use of a non-alcohol based disinfectant is recommended. Non-alcohol based products
contain the active ingredient referred to as quaternary ammonia compound. The
quaternary ammonia compound is specifically designed to disinfect rubber, plastic,
silicone and acrylic products which are commonly used in hearing evaluation devices.
Follow this process for cleaning and disinfection of the BERAphone
®
after use:
1. Clean residual gel from the electrodes and ear cushion using a tissue or a
disinfectant wipe.
2. Disinfect the electrodes, ear cushion and other components that made contact
with the baby or the baby’s bedding by wiping them with a fresh hospital-grade
disinfectant wipe.
a. Follow the directions for use and precautions on the disinfectant product.
b. If the disinfectant wipe is very wet, do not allow disinfectant to drip down into
the black, perforated speaker inside the BERAphone
®
ear cushion area.
3.
Disinfect any other components that touched the baby or the baby’s bedding
such as the cable, BERAphone
®
handle, etc.
4.
Allow the disinfectant to dry thoroughly according to the manufacturer’s
instructions for maximum efficiency before using it on the next patient.