Operation Manual easyScreen
8508935 Rev. 4
21
04/04/2019
Cleaning the OAE Probe Tip
In order to achieve accurate measurements it is important to make sure that the probe
system is kept clean at all times. Therefore check the probe for debris in the channels
after each use and clean them as needed. It is critical to remove cerumen from the probe
tip’s small acoustic channels. Therefore please follow the illustrated instructions below.
Never clean the probe tip while the tip is still attached to
the probe!
Figure 2
Use the Bridge & Implant Floss or the ProxySoft 3 in 1 Floss
for cleaning (Figure 2).
Discard floss after use.
Figure 3
Unscrew the probe cap by turning it in a counter clockwise
direction. (Figure 3).
Figure 4
Take the plastic probe tip out of the probe (Figure 4).
Figure 5
Insert the blue end of the floss from back to front through
one of the probe channels (Figure 5).
Figure 6
Pull the floss along its entire length through the channel
(Figure 6).
Proceed in the same way with all 4 probe channels. Use
the floss only once.
Figure 7
Use a thick or a thin floss depending on the size of the
channel (Figure 7).