Operation Manual easyScreen
8508935 Rev. 4
87
04/04/2019
The following table shows the protocols that can be installed in HearSIM
™
PROTOCOL FILE NAME
PARAMETERS
SENSITIVITY
ABRIS A00 CE-Chirp 35 dB nHL
CE-Chirp
®
stimulus of 35 dB nHL (default)
≥ 99.6%
ABRIS A01 CE-Chirp 30 dB nHL
CE-Chirp
®
stimulus of 30 dB nHL
≥ 99.6%
ABRIS A02 CE-Chirp 40 dB nHL
CE-Chirp
®
stimulus of 40 dB nHL
≥ 99.6%
ABRIS A03 CE-Chirp 45 dB nHL
CE-Chirp
®
stimulus of 45 dB nHL
≥ 99.6%
ABRIS A04 Click 35 dB nHL
Click stimulus of 35 dB nHL
≥ 99.6%
ABRIS A05 Click 30 dB nHL
Click stimulus of 30 dB nHL
≥ 99.6%
ABRIS A06 Click 40 dB nHL
Click stimulus of 40 dB nHL
≥ 99.6%
ABRIS A07 Click 45 dB nHL
Click stimulus of 45 dB nHL
≥ 99.6%
DPOAE D00 2-5 kHz, 3_4, SNR 6 dB F2 Frequencies: 5k, 4k, 3k, 2kHz (default)
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 4
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 6 dB
Artifact rejection level: 30 dB SPL
≥ 99.6%
DPOAE D01 1.5-4 kHz, 3_4, SNR 6 dB F2 Frequencies: 4k, 3k, 2k, 1.5k Hz
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 4
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 6 dB
Artifact rejection level: 30 dB SPL
≥ 99.6%
DPOAE D02 1.5-6 kHz, 3_5, SNR 6 dB F2 Frequencies: 6k, 4k, 3k, 2k, 1.5k Hz
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 5
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 6 dB
Artifact rejection level: 30 dB SPL
≥ 99.6%
DPOAE D05 1.5-6 kHz, 3_6, SNR 7 dB F2 Frequencies: 6k, 5k, 4k, 3k, 2k, 1.5k Hz
Level (L1/L2): 65/55 dB SPL
F2/F1 Ratio: 1.22
Max. test time: 60 s
# Freq. for pass: 3 / 6
Pass criteria for each frequency:
Min OAE level: -5 dB SPL
Min SNR: 7 dB
Artifact rejection level: 30 dB SPL
≥ 99.6%