9
There are three traverse lengths available. For measuring short surfaces such
as o-ring grooves, short lands and shoulders, use Switch Positions 1 or 3.
1RPLQDO
1XPEHURI&XWRIIV
7UDYHUVH/HQJWK
(YDOXDWLRQ/HQJWK
6ZLWFK3RVLWLRQ
.075"/2.0mm
.030"/0.8mm
1
.135"/3.5mm
.090"/2.4mm
3
.195"/5.0mm
.150"/4.0mm
5
Cutoff equals .030"/0.8mm.
7UDYHUVH/HQJWK
*DJH2SHUDWLRQ
The Pocket Surf gage will operate in any one of four different probe positions (see
illustration on next page). Selection of the proper probe position depends on the ap-
plication.
When changing probe positions, gently grasp the probe by its body; never handle the
probe by its skid and stylus end.
Prior to operating the gage, its calibration should be checked. Refer to the
&DOLEUDWLRQ
section for details.
Important:
During the measuring cycle, the probe stylus and skid should be in contact
with a surface, and the setup should be properly aligned (as outlined in the following
sections). Otherwise, any readings obtained are not valid and are not to be used for