13
The Pocket Surf gage is calibrated
using the Reference Specimen and the
EPL-1681 Riser Plate supplied with the
Pocket Surf Kit.
Set the Parameter Selection Switch to
“R
a
”. The other two parameters, R
max
and R
z
will automatically be calibrated
when R
a
is calibrated.
Set the Traverse Length Switch to
position 5 because a full 5 cutoffs are
required for proper calibration of the
instrument.
&DOLEUDWLRQ
Note:
When calibrating with a General Purpose
Probe, Groove Bottom Probe, the Transverse Chisel
Probe or a Small Bore Probe, the probe should be in
the 180° “Extended” position (as shown above). The
Parallel Chisel Probe must be in the 90° or 270° posi-
tion and the setup rearranged so that the probe will
traverse the Specimen in the direction of the arrow.
1. To elevate the Pocket Surf base to
the same height as the surface of
the Reference Specimen, place the
Pocket Surf unit on the Riser Plate
so that the two “V’s” on the bottom of
the Pocket Surf Unit sit in the shal-
low recesses of the Riser Plate. Align
the Reference Specimen so that the
gage probe will traverse across it in
the direction of the double-ended ar-
row. Take a reading from the center of
the specimen.
Riser Plate
Reference Specimen