To use the gage in the other probe positions — 90°, 180° or 270°, the Pocket Surf unit
and probe must be carefully aligned to the workpiece surface to be measured. The
base of the Pocket Surf unit must be at approximately the
VDPHHOHYDWLRQ
(within ±.100"/±2,5mm) as the surface being measured, and the axis of probe traverse
must be
SDUDOOHO
to the surface being measured.
With the probe positioned on the workpiece surface to be measured, carefully ad-
MXVW
the gage and workpiece setup so that the black, scribed Reference Line on the
Probe Mounting Block is
SDUDOOHO
to the bottom of the Pocket Surf housing and paral-
OHOWRWKHZRUNVXUIDFH7KLVHQVXUHVWKDWWKHSUREHVNLGDQGVW\OXVDUHÀXVKRQWKH
surface, even though (except for the Small Bore Probes) the probe body will be at an
angle to the surface.
2WKHU3UREH3RVLWLRQV
Parallel
Reference Line
Reference Line
Within ±.100"/±2.5mm
Within ±.100"/±2.5mm