11.6 

Suction circuit components 

The  suction  circuit  is  a  group  made  up  of  various 
pressurised  equipment  connected  through  hoses  and/or 
brazed piping. 
A  routine  maintenance  operation  is  the  replacement  of  the 
dryer  filter,  the  procedure  of  which  is  outlined  in  the  user 
manual. 
Safety  against  the  risk  due  to  pressure  of  the  refrigerant 
fluid inside the suction unit is ensured by a safety pressure 
switch  and  a  thermal  fuse  both  installed  during 
manufacturing. 
In  case  of  replacement  of  one  of  these  safety  devices  use 
new and original spare parts. 
The  assistance  centre  is  responsible  for  workmanlike 
assembly  and  for  checking  before  new  commissioning 
according  to  the  prescriptions  provided  for  by  the 
regulations  in  force  in  the  country  of  use;  in  particular,  the 
assistance  centre  will  have  to  guarantee  that  the 
components  ensure  a  perfect  tightness  and that  the safety 
components  are  installed  appropriately  so  as  to  ensure 
operation is not jeopardized. 

 

Fig. 25:  Suction circuit position 

 

PRESSURE SWITCH ACB-2UB949W o ACB-2UB1145W 
13/18bar 1/4SAE Category IV (Dir. 2014/68/EU) 

 

12 Non

-permanent junction 

 
Here  below  some  instructions  on  the  use  of  Loctite®  577  on  the 
thread of some non-permanent junctions in the pneumatic circuit of 
the ACX  stations. 
The  product  polymerizes  when  air  is  absent  and  when  in  contact 
with

 

metal

its aim is preventing slackening caused by impacts and 

vibrations. Yet, after application and fixing onto the junction/thread, 
you must tighten and then wait several hours, in relation to the type 
of  coupling  between  metals,  before  the  product  can  get  its 
maximum performance in terms of hold/resistance. 
For  one’s  own  and  other  people’s  safety,  the  operator/user  of  the 
Loctite®  577  product  must  consider  that  it  must  be  employed  in 
well-ventilated  premises  and  that  protecting  eyes  and  hands  with 
suitable  PPE  (personal  protective  equipment)  is  required.  For 
compulsory  safety  provisions  to  be  followed  when  using  the 
product, the user shall refer to the safety data sheet of Loctite® 577 
manufacturer. 
For  further  details  on  the  product  characteristics/performance  and 
application  instructions,  please  refer  to  the  technical  data  sheet  of 
Loctite® 577 manufacturer. 
 
Generally  the  points  of  insertion  of  Loctite®  577  only  in  one  spot 
with  anti-rotation  purpose  are  on  connection/fittings  where  there  is 
not an O-ring/seal, before to apply it clean well the surfaces. 

 

13 Electrical connections 

The  electrical  connections  are  described  on  the  following 
APPENDIX: 
 
APPENDIX 3 – ACX380/480 - ELECTRICAL CONNECTION – LOGIC 
ELECTRONIC BOARD 
 
APPENDIX  4  –  ACX380/480  -  ELECTRICAL  CONNECTION  – 
POWER ELECTRONIC BOARD AND POWER SUPPLY  
 
APPENDIX  5  –  ACX380/480  -  ELECTRICAL  CONNECTION  – 
STATUS AND WARNING INDICATOR LIGHT ELECTRONIC BOARD

  

Pressure Safety Switch 

(on the red circle) 

 

Содержание ACX380

Страница 1: ...MAHLE ACX380 480 EN Service Manual A C Service Equipment...

Страница 2: ...rigerant fluid circuit 17 11 1 Refrigerant fluid circuit 17 11 2 Operational cycles 18 11 3 Receiving tank refrigerant cylinder 18 11 4 Manifold charts 18 11 5 NITROGEN TEST and FORMING GAS LEAK TEST...

Страница 3: ...ing of the planet The release of refrigerant gas R134a into the atmosphere is forbidden by laws that were enacted within the framework of the Kyoto protocol For information only we will mention in par...

Страница 4: ...h an original spare part or equivalent by a MAHLE centre Before opening the service station extract completely the supply cable from the plug or you can get an electric shock Do not tamper with or byp...

Страница 5: ...f instructed or certified for the equipment open the equipment The equipment parts which can cause electrocution shut off power to the equipment before servicing repairing it 2 3 Safety devices The A...

Страница 6: ...by the vehicle s manufacturer Charge filling of refrigerant into the A C system in the amount specified by the manufacturer System flushing Cleaning phase for the removal of possible polluting substa...

Страница 7: ...ant recovery and recharging Vacuum generation Flushing 4 2 Scope of delivery Description Service hose high pressure Service hose low pressure Quick release coupling high pressure Quick release couplin...

Страница 8: ...are available on the display Icon Description Device connected via WiFi The residual refrigerant weight in kg and with visual bar to enable disable the bar by vertically swiping your finger from top t...

Страница 9: ...menu The Main menu of the graphical user interface allows to select the following functions Vehicle selection OneClick Cycles Additional Functions Maintenance Setup Secured Service not available for u...

Страница 10: ...014 68 EU IV Pneumatic fittings Net length of external HP and LP hoses 4 5 m HP and LP Analog 80 mm pressure gauges pulse free 1 0 class Display Touch screen 7 TFT colours Keypad Touch screen Software...

Страница 11: ...for correct handling shall be ensured as provided for by accident prevention provisions POSITIONING Place the unit in a stable place The location must be well ventilated with a good rate of change of...

Страница 12: ...coupler Tanks without plunger have only one valve so they must be turned upside down to transfer the liquid refrigerant The LP gauge indicates the pressure inside the external tank After some minutes...

Страница 13: ...side the auto vehicle A C system To fill the UV dye bottle Fig 1 Pos 9 it has to be extracted from its housing by means of the quick coupler on the top of the bottle slightly press downward the couple...

Страница 14: ...the garage data to be printed at the bottom of the cycle report LANGUAGE by selecting this entry any language present in the database may be set In case you choose a language with unintelligible char...

Страница 15: ...ot connected to the bottle station accessories kit when the next message is displayed Fig 16 LP quick coupler opening Turn the LP quick coupler clockwise to open it and press enter OK so that air gets...

Страница 16: ...pening Turn the LP quick coupler clockwise to open it and press enter OK so that air gets into the LP hose Wait for the message to connect the reference gauge contained in the service kit and by rotat...

Страница 17: ...far from the value of the certified reference weight the result is negative in this case is not possible to update calibration It is necessary to carry out the 3 points Gas weight calibration mention...

Страница 18: ...ation is not jeopardized VESSEL 20l Category III Dir 2014 68 EU Fig 20 Refrigerant internal bottle SAFETY VALVE GAS AIRTEK VS14NPT20HNBRPED4 PS 20 bar R 1 4 NPT Category IV Dir 2014 68 EU 11 4 Manifol...

Страница 19: ...n MAHLE Fig 22 Manifold circuit Fig 23 Manifold valves Fig 24 Manifold with pressure sensors installed V A I R L1 V A I L2 R T1 T1 L1 R A V PS TP PS HP PS LP One way valve L2 V A I L2 R T1 L1 R A V PS...

Страница 20: ...LD The unit has a second manifold dedicated to nitrogen test the identification letters of the sectioning valves are highlighted is reported as follows T2 nitrogen input N IN X nitrogen output N OUT X...

Страница 21: ...izes when air is absent and when in contact with metal its aim is preventing slackening caused by impacts and vibrations Yet after application and fixing onto the junction thread you must tighten and...

Страница 22: ...and scroll to MAINTENANCE and then select Fig 26 Maintenance screens Only for model ACX380 REFRIGERANT TYPE CHANGE With this function allows the service assistance technician to change the Refrigeran...

Страница 23: ...he main menu select MAINTENANCE and press PRESSURES ZERO This function allows to determine and store the at pressure value We recommend running this procedure every time the A C service station will b...

Страница 24: ...er to zero and to start the filter replacement procedure Insert the code of the new filter Now you can replace the filter Required tools 1 flat screwdriver 1 cross screwdriver 1 regular or torque Hex...

Страница 25: ...and press CALIBRATION CHECK This function allows to check the status of gas weight scale measure with a reference weight from 100 g to 10000 g maximum not included in the scope of delivery When reque...

Страница 26: ...ovide for a the bodies in charge after the results of the previous checks performed of commissioning or period requalification do not establish different The checks must be performed by the bodies in...

Страница 27: ...install the specific components for use of R1234yf He will also carry nfigurations and verifications required by the the MAINTENANCE 14 17 Replacing white sample filter Refrigerant identification unit...

Страница 28: ...ronic power board Replacement of power supply board 15 2 Extraordinary maintenance 15 2 1 Replacement of component connected to refrigerant circuit The components pressurised equipment constituting co...

Страница 29: ...tery Evident damage to board components and circuits For the replacement of the board it is recommended to position the new board alongside the one to be replaced Disconnect and reconnect the cables i...

Страница 30: ...C SERVICE STATION ACX380 480 Blue LP quick coupler and red HP quick coupler Active Oil Protection kit 2 pcs New Oil 1 UV dye rts are available through the Service Centers USING NON ORIGINAL UNAPPROVE...

Страница 31: ...en 31 ACX380 480 APPENDIX 1 REFRIGERANT FLUID CIRCUIT ACX380 480 H...

Страница 32: ...ACX380 480 32 en APPENDIX 2 REFRIGERANT FLUID CIRCUIT ACX380 480 OPERATIONAL CYCLES Uncondensable gases H...

Страница 33: ...en 33 ACX380 480 Recovery cycle H...

Страница 34: ...ACX380 480 34 en Vacuum H...

Страница 35: ...en 35 ACX380 480 Recharge H...

Страница 36: ...load cell dye 1 10 LC New Oil1 input load cell new oil 1 11 PS Tank input pressure sensor tank 12 LC Ex Oil input load cell 13 PS LP input pressure sensor LP 14 PS HP input pressure sensor HP 15 TS Ta...

Страница 37: ...7 ACX380 480 APPENDIX 4 ACX380 480 ELECTRICAL CONNECTION POWER ELECTRONIC BOARD AND POWER SUPPLY POWER ELECTRONIC BOARD POWER SUPPLY BOARD 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 24 2...

Страница 38: ...ng 9 OUT H Heat exchanger fan 10 OUT I HP LP solenoid interconnection valve 11 OUT N Solenoid valve for draining of tank s uncondensable gases 12 OUT V Vacuum pump 13 OUT V Vacuum solenoid valve coupl...

Страница 39: ...n 39 ACX380 480 APPENDIX 5 ACX380 480 ELECTRICAL CONNECTION STATUS AND WARNING INDICATOR LIGHT ELECTRONIC BOARD Item label Description Connected Unused x 1 SER I O serial I O interface to power board...

Страница 40: ...compressor check for leaks from valves R or check compressor functioning Pressure sensor not Altitude compensation and or full pressure calibration Mistake during procedure Altitude compensation and...

Страница 41: ...acuum pump oil When the filter capacity is the filter capacity is finished Change filter with new one During the setting of the inner tank charge amount Amount required greater than that available in...

Страница 42: ...d or damaged To check electrical connections or change with new spare part During vacuum bottle filling or leak test phase in vacuum The vehicle A C system is pressurized Recover the refrigerant gas f...

Страница 43: ...leak test Empty the vehicle follow the procedure guided by the displayed messages A048 HP LEAKAGE During the gas recovery and hoses emptying phase At the end of the gas charging in the E3 Connect qui...

Страница 44: ...nalisys not possible Check with virgin refrigerant bottle if the refrigerant gas identifier is working properly If yes and the refrigerant of the car is not R1234yf empty in a safety way and in compli...

Страница 45: ...MAHLE Aftermarket Italy S p A Via Quasimodo 4 a 43126 Par ma ITALY 2019 10 28...

Отзывы: