Magtrol Model 6530 Power Analyzer
Appendix A: Schematics
appendices
Magtrol 7500 Series Power Analyzer
Appendix B: Compatible Command Sets
Command Code
function
explanation
MAm1,m2<terminator> Sets the RMS or DC
measurement mode of amps
phases.
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates measurement mode.
Values for
m2
are:
0 = RMS (AC + DC)
1 = DC
MVm1,m2<terminator> Sets the RMS or DC
measurement mode of voltage
phases.
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates measurement mode.
Values for
m2
are:
0 = RMS (AC + DC)
1 = DC
MCm<terminator>
Sets the continuous or cyclic
measurement mode filter.
"m" indicates whether the measurement
mode is continuous or cyclic.
Values for
m
are:
0 = continuous
1 = cycle-by-cycle
MFm<terminator>
Sets the AC and DC
measurement mode filter.
"m" indicates the measurement filter
low-pass frequency (response time).
Values for
m
are:
0 = 1 Hz
1 = 2 Hz
2 = 5 Hz
3 = 10 Hz
4 = 20 Hz
5 = 50 Hz
6 = 100 Hz
PC<terminator>
Clears all peak hold values
(volt, amp and watt).
---
RAm1,m2<terminator> Sets current range and input
source.
"m1" indicates the input phase.
Values for
m1
are:
0 = all phases
1 = phase 1
2 = phase 2
3 = phase 3
"m2" indicates current range.
Values for
m2
are:
0 = 20 A
1 = 10 A
2 = 5 A
3 = 1 A
4 = 1 V external sensor input range
5 = 500 mV external sensor input range
6 = 250 mV external sensor input range
7 = 50 mV external sensor input range