Magtrol Model 6530 Three-Phase Power Analyzer
Chapter 5 – Computer Controlled Operation
OPERA
TION
Magtrol 7500 Series Power Analyzer
Command Code
function
explanation
CONF:MEASFILTER
<?|filter>
Sets the AC and DC
measurement filter.
<filter> indicates the low pass frequency
filter used.
Valid values for <filter> are:
? returns the current filter choice
0
1 Hz
1
2 Hz
2
5 Hz
3
10 Hz
4
20 Hz
5
50 Hz
6
100 Hz
CONF:IRANGE
<phase>,<?|range>
Sets the input source and
acceptable current range.
<phase> indicates the input phase
Valid input s for <phase> are:
0
all phases
1
phase 1
2
phase 2
3
phase 3
<range> indicates current range.
Valid values for <range> are:
? returns the phases current setting
0
20 A
1
10 A
2
5 A
3
1 A
4
1 V (ext sensor input range)
5
500 mV (ext sensor input range)
6
250 mV (ext sensor input range)
7
50 mV (ext sensor input range)
CONF:ISCALE
<phase>,<?|scale>
Sets the amps scaling
constant.
<phase> indicates the input phase
Valid input s for <phase> are:
0
all phases
1
phase 1
2
phase 2
3
phase 3
<scale> indicates the amps scaling constant
in A/A.
Valid values for <scale> are
? to return the current scale setting
floating point numbers within the following
range:
0.01 < scale < 10000.0
When <scale> is set to 0 scaling mode will
be cleared
CONF:VRANGE
<phase>,<?|range>
Sets the voltage input
range for a certain phase
<phase> indicates the input phase
Valid input s for <phase> are:
0
all phases
1
phase 1
2
phase 2
3
phase 3
<range> indicates voltage range.
Valid values for <range> are:
? returns current voltage range setting
0
600 V
1
300 V
2
150 V
3
30 V