38
Magtrol Model 6530 Three-Phase Power Analyzer
Chapter 4 – Operating Principles
THEOR
Y
Magtrol 7500 Series Power Analyzer
4.2.2
Dc D
etaiLs
4.2.2.1
Interrupt Driven
The power analyzer will automatically update its data every 4.469 microseconds by completing the
following sequence of events:
• The volts input is read and checked for an over range condition. The offset is then added
to the reading and the reading is negated and saved.
• The amps input is read and checked for an over range condition. The offset is then added
to the reading and the reading is negated and saved.
• The volt reading is checked to see if it is the peak and if so, is stored.
• The amp reading is checked to see if it is the peak and if so, is stored.
• The volt reading is summed.
• The amp reading is summed.
• The volt and amp reading is multiplied.
• The multiplied value is summed.
Note:
After 256 readings are summed the sums are transferred to a 64-bit
V register, a 64-bit I register, and a 64-bit VI register. A sample
ready flag is set.
4.3
measurement modes
4.3.1
p
eak
The peak value is defined as the highest value that occurs in the cycle of a repetitive waveform.
The graph below shows one cycle of a repetitive waveform. The peak value is shown. The absolute
value of the reading is used to determine the peak value.