36
Magtrol Model 6530 Three-Phase Power Analyzer
Chapter 4 – Operating Principles
THEOR
Y
Magtrol 7500 Series Power Analyzer
4.2.1
ac D
etaiLs
4.2.1.1
Interrupt Driven
The power analyzer will automatically update its data every 4.469 microseconds by completing the
following sequence of events:
• The volts input is read and checked for an over range condition. The offset is then added
to the reading and the reading is negated and saved.
• The amps input is read and checked for an over range condition. The offset is then added
to the reading and the reading is negated and saved.
• The volt reading is checked to see if it is the peak and if so is stored.
• The amp reading is checked to see if it is the peak and if so is stored.
• The volt reading is squared and the squared reading is summed.
• The volt and amp reading is multiplied and the multiplied value is summed.
• The amp reading is squared and summed.
Note:
After 256 readings are summed, the sums are transferred to a 64-bit
V² register, a 64-bit VI register, and a 64-bit I² register. A sample
ready flag is set.