Magtrol Model 6530 Three-Phase Power Analyzer
Chapter 5 – Computer Controlled Operation
OPERA
TION
Magtrol 7500 Series Power Analyzer
Command Code
function
explanation
MEAS:WATTS
<phase>,<type>
Returns the power
measurement at the
current instant.
<phase> indicates input phase.
Valid values for <phase> are:
0
sum of watts phases
1
phase 1
2
phase 2
3 phase 3
<type> indicates the type of measurement.
Valid values for <type> are:
0
watts
1
VA
2
Power factor
3 Watts peak hold
5.7.3
|conF|
iguration
c
oMManDs
NOTE:
The current configuration can be read by replacing the parameter
with a question mark “?”Example: CONF:FILTER ? will return
the current filter setting
Command Code
function
explanation
CONF:AMPS
<phase>,<?|range>
Sets auto or manual range
mode for the amp phases.
<phase> indicates the input phase
Valid inputs for <phase> are:
0
all phases
1
phase 1
2
phase 2
3
phase 3
Valid inputs for <range> are:
? returns the phases current setting
0
manual range
1
auto range
CONF:VOLTS
<phase>,<?|range>
Sets auto or manual range
mode for the volts phases.
<phase> indicates the input phase
Valid inputs for <phase> are:
0
all phases
1
phase 1
2
phase 2
3 phase 3
Valid inputs for <range> are:
? returns the phases current setting
0
manual range
1
auto range