7. Trouble Information
Check the following items and take the proper remedial actions when the status indicators of the
interpolator unit indicate an error or an alarm.
Symptom
Cause
Remedial action
Th
e STATUS indicator lights
up or blinks red.
Self-calibration failed.
See section 3-1 “Installation” and
section 4-1 “Measurement when
Performing Self-calibration,” and
perform installation again.
A SENSOR (1 to 6) indicator
lights up red.
A level alarm or speed alarm has
occurred.
In case of a level alarm:
Th
e sensor signal level has dropped.
a
Dropped due to dirtiness
See section 6-1 “Cleaning” and clean
the scale unit.
b
Th
e installation position may not
satisfy the tolerance range.
See section 3-1 “Installation” and
section 8-2 “Installation Error Range
and Environmental and Rotational
Accuracy Conditions,” and check the
installation position.
In case of a speed alarm:
Th
e maximum rotary response speed
was exceeded.
Use at the maximum rotary response
speed or less.
Th
e POWER indicator is off .
Th
e power is not turned on.
Check that power supply (+24 V) is
input. Check whether the fuse has
melted.
If performing the above actions does not resolve the trouble, contact our service department.
Содержание Laserscale SET-HD100
Страница 3: ...SET HD100 J 1...
Страница 5: ...SET HD100 J 3...
Страница 8: ...SET HD100 6 J 1 2 12 2 1...
Страница 10: ...ii J SET HD100 9 22 9 1 22 9 2 24...
Страница 11: ...J 1 SET HD100 1 1 1 LS ARC 1 2 360 0 2 1 9 10 5 USB 1 3 DC20 24 V 5 8 A AC USB...
Страница 12: ...2 J SET HD100 1 4 BE300 1 BH300 1 BD350 1 CS03 05 1 1 1 1 2 1 1 1 1 1 1...
Страница 13: ...J 3 SET HD100 2 2 1 2 1 1 OFF ON 2 2 1...
Страница 16: ...6 J SET HD100 3 50 mm 9 3 1 3 1 1 100 1000 3 1 2...
Страница 17: ...J 7 SET HD100 3 1 3 1 2 1 503 0 02 mm 1 2...
Страница 19: ...J 9 SET HD100 3 2 1 0 6 0 06 N m 4 4 4 5...
Страница 20: ...10 J SET HD100 6 5 4 1 0 0 1 N m 3 2 OFF ON 3 2 1 DC 24 V AC BP 24 USB USB2 0 Type B 3 2 2 1 OFF 2 DC 3 USB...
Страница 23: ...J 13 SET HD100 5 0 3 0 4 N m...
Страница 24: ...14 J SET HD100 3 1 2 BH300 BE300 BD350...
Страница 25: ...J 15 SET HD100 CS03 05 4 5...
Страница 26: ...16 J SET HD100 6 7...
Страница 27: ...J 17 SET HD100 6 6 1 6 2 1 1...
Страница 28: ...18 J SET HD100 7 STATUS 3 1 4 1 SENSOR 1 6 6 1 3 1 8 2 POWER 24 V...
Страница 34: ...24 J SET HD100 9 2 110 118 210 298 mm...
Страница 58: ...SET HD100 16 E Carry case Bands Lock Secure with the bands 6 Close and lock the lid Storage complete 7...
Страница 66: ...SET HD100 24 E 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm...
Страница 67: ......
Страница 69: ...SET HD100 1...