Mount the scale unit
While maintaining the eccentricity adjustment performed in Step 2, fully tighten the screws semi-
tightened in Step 1.
<Specifi cation>
Tightening torque: 0.6 ±0.06 N·m
* When the reproducibility is insuffi
cient, warping may have occurred due to tightening. After tightening the screws, it is
recommended to loosen and then retighten the screws in the order that the screws were tightened.
4
Temporarily mount the head unit
Set the head unit on the head unit mounting surface with the screw holes aligned, and semi-tighten
the screws. Use screws procured by the customer.
Head unit
Screw × 4
Head unit
mounting base
<When mounting using the head unit upper reference surface>
Head unit
Screw × 4
Head unit mounting base
<When mounting using the head unit lower reference surface>
5
Axes alignment of the head unit and scale unit
Insert the supplied axis alignment jig to the inner diameter parts of the head unit and scale unit.
Adjust position of head unit so that center axis of head unit get into alignment with center axis of scale
unit.
Notes
• Take care not to damage the inner diameter part of the scale unit when inserting the axis alignment
jig.
• When mounting the scale unit and head unit facing sideways, take care so that the head unit or
other items do not fall when working in the temporarily mounted condition.
Содержание Laserscale SET-HD100
Страница 3: ...SET HD100 J 1...
Страница 5: ...SET HD100 J 3...
Страница 8: ...SET HD100 6 J 1 2 12 2 1...
Страница 10: ...ii J SET HD100 9 22 9 1 22 9 2 24...
Страница 11: ...J 1 SET HD100 1 1 1 LS ARC 1 2 360 0 2 1 9 10 5 USB 1 3 DC20 24 V 5 8 A AC USB...
Страница 12: ...2 J SET HD100 1 4 BE300 1 BH300 1 BD350 1 CS03 05 1 1 1 1 2 1 1 1 1 1 1...
Страница 13: ...J 3 SET HD100 2 2 1 2 1 1 OFF ON 2 2 1...
Страница 16: ...6 J SET HD100 3 50 mm 9 3 1 3 1 1 100 1000 3 1 2...
Страница 17: ...J 7 SET HD100 3 1 3 1 2 1 503 0 02 mm 1 2...
Страница 19: ...J 9 SET HD100 3 2 1 0 6 0 06 N m 4 4 4 5...
Страница 20: ...10 J SET HD100 6 5 4 1 0 0 1 N m 3 2 OFF ON 3 2 1 DC 24 V AC BP 24 USB USB2 0 Type B 3 2 2 1 OFF 2 DC 3 USB...
Страница 23: ...J 13 SET HD100 5 0 3 0 4 N m...
Страница 24: ...14 J SET HD100 3 1 2 BH300 BE300 BD350...
Страница 25: ...J 15 SET HD100 CS03 05 4 5...
Страница 26: ...16 J SET HD100 6 7...
Страница 27: ...J 17 SET HD100 6 6 1 6 2 1 1...
Страница 28: ...18 J SET HD100 7 STATUS 3 1 4 1 SENSOR 1 6 6 1 3 1 8 2 POWER 24 V...
Страница 34: ...24 J SET HD100 9 2 110 118 210 298 mm...
Страница 58: ...SET HD100 16 E Carry case Bands Lock Secure with the bands 6 Close and lock the lid Storage complete 7...
Страница 66: ...SET HD100 24 E 9 2 Interpolator Unit 210 110 118 298 Serial No Unit mm...
Страница 67: ......
Страница 69: ...SET HD100 1...