Special methods 

30 

 
 

NANOCOLOR

®

 

UV

/

VIS

 II and 

NANOCOLOR

®

 

VIS

 II   V 1.1 / 10.2017 

 

Figure 31: Entering the properties of a special method in the single wavelength mode  

 

 

Figure 32: Entering the properties of a special method in the multi-wavelength mode 

 

Method  name:

  Enter  a  method  name  via  the  text  keypad.  The  name  should  not  exceed  20 

characters in length. Confirm the entry with Enter. 

 

Method  number:

  The  method  number  is  a  specific  number  with  which  the  method  can  be 

retrieved  later  from  the  list  box  in  the  "User  methods"  menu.  Enter  a  method  number 
between  1  and  99  using  the  numeric  keypad.  The  lowest  available  number  will  appear 
automatically.  Confirm  the  entry  with  OK.  Should  the  chosen  method  number  not  be 
available,  the  message  "This  method  number  is  already  in  use"  will  appear.  Either  select 
another method number or delete the method having that number. 

 

Symbol:

 Enter a symbol via the text keypad which will be shown after the unit when the result 

is displayed. Confirm the entry with Enter. No more than 10 characters are allowed. 

 

 

Содержание NANOCOLOR UV/VIS II

Страница 1: ......

Страница 2: ...iew 10 3 2 Back view 11 4 Operating instruction 11 4 1 Connectivity 11 4 2 Turning on the instrument 11 4 3 Operation and user guidance 12 4 3 1 Operation of the touch screen 12 4 3 2 Task and status...

Страница 3: ...5 Yellowness Index 49 5 6 6 6 Hazen APHA PtCo color index 49 5 6 7 Visual color scales 50 5 6 7 1 Iodine color index 50 5 6 7 2 Ph Eur color index 50 5 6 8 Distinction of the color index according to...

Страница 4: ...additions 80 6 6 6 IQC counter 82 6 6 7 IQC memory 82 6 6 8 IQC Card 4 84 6 7 User accounts 86 6 8 Pictograms 88 6 9 Data export 89 6 9 1 LIMS 89 6 9 2 ACRON 90 6 10 Tools 90 6 10 1 Timer 91 6 10 2 L...

Страница 5: ...Content iv NANOCOLOR UV VIS II and NANOCOLOR VIS II V 1 1 10 2017 10 3 Spare parts accessories and consumables 98 10 4 Contact 99 10 5 Warranty liability and complaints 99 10 6 Waste disposal 99...

Страница 6: ...armonization legislation of the European Community listed below European Directive 2011 65 EU on the restriction of the use of certain hazardous substances in electrical and electronic equipment RoHS...

Страница 7: ...em Monochromator Wavelength range 190 nm 1100 nm 320 nm 1100 nm Wavelength accuracy 1 nm Wavelength resolution 0 1 nm Wavelength calibration Automatic Wavelength selection Automatic barcode manual Sca...

Страница 8: ...king with chemicals can be hazardous Handling chemicals without protection can lead to serious injuries When working with the NANOCOLOR spectrophotometers wear the required personal protective equipme...

Страница 9: ...terium lamp generates ultraviolet radiation which can harm your eyes Never look directly into the light source without suitable UV protection glasses Also protect your skin against direct UV light The...

Страница 10: ...slot cover Protective covering Power plug only for NANOCOLOR UV VIS II Power supply with country specific adapters only for NANOCOLOR VIS II USB connection cable Quickstart guide Calibration cuvette M...

Страница 11: ...break nor other damages Otherwise there is an imminent danger of an electric shock 4 1 Connectivity Next to the electrical connection four more types of interfaces are available within the NANOCOLOR U...

Страница 12: ...capacitive touch screen PCAP The instrument s cover glass is non reflecting It can be easily and conveniently cleaned with the microfiber cloth included in your package or with a soft cotton cloth Fi...

Страница 13: ...numbers or special characters Figure 10 Text keypad II Pressing the key on the special character layout will switch to another keypad of special characters Via the text keypad this key is used to capi...

Страница 14: ...ease consider that some menus cannot be retrieved by pressing the icon e g measurement menu as the process has been ended by leaving the respective menu Feature icon By tapping this icon one can activ...

Страница 15: ...ng a checkbox with a checkmark deactivates the box and removes the checkmark 4 3 5 List functions When editing text or numeric fields it is possible that a selection window with a list will be opened...

Страница 16: ...firming with opens the measurement window Inserting the zero solution and pressing will initiate the zero measurement After the successful measurement of the zero solution the sample solution needs to...

Страница 17: ...figure 18 Figure 18 Result window of the absorbance measurement The absorbance of the individual wavelengths can be displayed by selecting the appropriate wavelength from the list on the left Use the...

Страница 18: ...zero measurement After the successful measurement of the zero solution the sample solution needs to be inserted Pressing starts the measurement process After the measurement has been completed the me...

Страница 19: ...con from the sample information bar the procedure is similar This time the icon must be removed from the blue flashing sample information bar with a wiping motion after holding it down The following s...

Страница 20: ...s still within the measuring range but outside of the 20 80 measuring range If the bar is colored red the measured value is out of range The Correction measurement option can be used when a correction...

Страница 21: ...t box 5 2 2 Standard tests MACHEREY NAGEL standard tests can be selected in different ways It is however not possible to select a standard test by barcode identification The appropriate measurement me...

Страница 22: ...ys requires determination of a correction value Color and turbidity both cause increased light absorption increased absorbance which leads to wrong results Determination of correction values requires...

Страница 23: ...0 Almost all colors and turbidities are destroyed under test conditions and do not interfere COD resistant colors and turbidities cause deviating results which cannot be circumvented 0 31 Cyanide 08 P...

Страница 24: ...ube add 4 0 mL sample solution close mix 0 70 POC 200 Proceed as described in the instructions for test 0 70 but instead of 1 0 mL R2 add 1 0 mL distilled water 0 71 Nickel 4 Proceed as described in t...

Страница 25: ...lution consider dilution mix add 2 drops R3 close and mix 8 59 Methanol 15 Open methanol test tube add 3 0 mL R1 and 1 5 mL sample solution consider dilution mix add 2 drops R3 close and mix 8 71 Pero...

Страница 26: ...s MEBAK 2 23 Thiobarbituric acid value MEBAK 2 4 acids MEBAK 2 18 2 Iso acids MEBAK 2 18 2 Cyclamate MEBAK 3 11 4 Iron MEBAK 2 29 3 Free amino nitrogen FAN MEBAK 2 8 4 1 Total carotenoids acc to Weser...

Страница 27: ...Figure 26 Options for user method creation 5 4 2 1 Lists Pressing the icon opens the list containing all of the special methods that have already been created by the user Besides the scroll function...

Страница 28: ...er pressing the icon a list box will open showing the methods on the mass storage device available for import see Figure 28 Tapping on the desired entry will import the special method into the list of...

Страница 29: ...m the list and confirm with 5 4 2 2 Design The design menu allows you to create special methods based on previously established method information such as factors and wavelengths of the measurement In...

Страница 30: ...specific number with which the method can be retrieved later from the list box in the User methods menu Enter a method number between 1 and 99 using the numeric keypad The lowest available number wil...

Страница 31: ...le Calculation formula multi wavelength measurements In this section the wavelengths and coefficients given in the selected calculation formula are defined The number of wavelengths is determined by t...

Страница 32: ...en a numeric keypad for entering numerical values Negative factors can be depicted by entering a minus sign before the numeric value Please note The formula has the format y F0 F1 E F2 E2 F3 E3 F4 E4...

Страница 33: ...he statistical certainty level of the calibration function a multiple determination can also be performed for the respective concentrations To do so enter the desired value of the concentration The ab...

Страница 34: ...1 E2 as absorbance value of wavelength 2 K1 as a multiplication factor of the wavelength 1 K2 as a multiplication factor of the wavelength 2 etc To include a subtraction step in one of the formulas th...

Страница 35: ...to the graph and can be included or excluded by selecting or deselecting the checkbox in front of the pair of values in the analysis Following the deselection or selection of data pairs the calibrati...

Страница 36: ...ng and tapping the entry Statistical data see Figure 40 Figure 40 View of the statistical data of the user defined special method When creating calibration curves outliers can occur due to errors in t...

Страница 37: ...laced and the calibration function and statistical data will be re calculated To increase the accuracy of a method where there is a linear relationship between concentration and absorbance each concen...

Страница 38: ...e entry Print In the case of special methods based on a calibration as described in Section 5 4 2 3 the measured absorbance and concentration pairs can be viewed and edited via the entry Edit calibrat...

Страница 39: ...be entered using the icons located below the result window Press the icon to view other available options Via the Properties command new sample information can be added below the measurement menu see...

Страница 40: ...ne can decide via option buttons to perform either a color measurement or a color comparison against a reference as far as a color reference is defined via the color analysis menu already yet figure 4...

Страница 41: ...assigned with a serial number To analyze the color measurement use the Analysis command or select the result from the memory and press Pressing the Do not save command prevents the measurement result...

Страница 42: ...hree dimensional coordinate system containing all colors that the human eye can perceive Every color clearly corresponds to a point in the coordinate system The position of a color in the color space...

Страница 43: ...e the color of a batch with the color of a product specification the procedure of color comparison described in this section can be used The color references function was conceived for these kinds of...

Страница 44: ...e list in the middle of the color analysis menu as well as a color reference from the list of color references must be choosen The calculated result is expressed in the result view as a DE value Figur...

Страница 45: ...e shows the result of the measurement in the result view screen figure XX The device automatically evaluates the measurement The measured DE value is compared against the accepted DE value of the colo...

Страница 46: ...nalysis mode in the color analysis menu Mark the color measurements which are supposed to be compared The device automatically calculated a maximal DE value which is display in the result view Figure...

Страница 47: ...gar color beer color according to EBC and ASBC yellowness index Hess Ives color unit coloring according to Ph Eur Gardner color index ADMI color index ASTM color index Saybolt color index Klett color...

Страница 48: ...ightest value to 18 darkest value and is measured in 10 mm cuvettes The Gardner color indexes are defined by the CIE chromaticity coordinates x and y All of the color indexes are defined by a calibrat...

Страница 49: ...mg L Pt but it could also be colored red green or blue The ADMI values are not obtained by comparison with the PtCo calibration but rather calculated based on the standard AWWA 2120F through a 4th deg...

Страница 50: ...color standards for red cobalt chloride HCl blue copper chloride HCl and yellow iron III chloride HCl These are used to mix five color schemes for the colors red brown brown yellow yellow and yellow g...

Страница 51: ...urement value over a range of about 50 nm Klett photometers are still used today in bioanalysis for monitoring the growth of cell cultures However it is not the absorption that is measured but rather...

Страница 52: ...asuring takes longer than with other colorimeters A color measurement ZERO or sample takes about 30 seconds at an integration from 360 nm to 830 nm 5 6 13 Combinations of illuminants observers and cuv...

Страница 53: ...able 3 Allowable combinations of illuminants observers and cuvette types 5 7 Test number Besides opening the measurement menu of the cuvette tests via barcode detection it can also be opened by select...

Страница 54: ...nu Press the icon to open the main menu Now you can choose from the options shown in Figure 59 Figure 59 Main menu 6 1 Settings In this menu general settings for the instrument can be made or options...

Страница 55: ...applied Changing the regional settings will affect the display of the measurement result This can change the way the thousands separator and the decimal place are depicted Example DE Germany 1 567 22...

Страница 56: ...the date and time display can be seen in the lower left hand corner of the pop up window Figure 63 Setting the date and time 6 1 4 Memory settings Press the icon to view the memory settings Besides an...

Страница 57: ...ent results memory is only available in the Open Mode see also Section 6 6 2 The setting will be saved upon exiting the menu 6 1 5 Acoustic signal Press the icon to adjust and test the sound level of...

Страница 58: ...contain all information about a measurement In case the checkbox is not set the measurement results on the printout will only contain the most important information Setting the checkbox Inverted print...

Страница 59: ...Figure 68 will activate the function and a reaction time flag will appear in the status bar The setting will be saved upon exiting the menu Figure 68 Acivate Reaction time 6 1 9 Lamp control Press th...

Страница 60: ...ansmission in tube tests The result of the turbidity measurement is stored together with the actual measurement result Press the icon to open a window for turning on this function Mark the checkbox to...

Страница 61: ...be applied To connect the sipper pump to the spectrophotometer please consult Section 6 5 1 and observe the instructions in the sipper manual The settings will be saved upon exiting the menu Figure 71...

Страница 62: ...r to continue the process Confirming with OK opens a list of backup files available on the internal SDHC card The file names are marked with the date and time Select a backup file in order to import i...

Страница 63: ...ou set the time for the display to switch off when not in use in order to save energy Simply tap on the black display in the energy saving mode to switch it back on and start using the instrument The...

Страница 64: ...re version and the number of system starts are listed Please have this information ready when contacting the MACHEREY NAGEL technical support Figure 76 System info 6 2 2 System check Press the icon to...

Страница 65: ...icons and select the required update file from the list Wait until a message appears indicating the termination of the update process This process can take up to a few minutes Restart your device whe...

Страница 66: ...ibration will continue automatically When calibration is complete the date of calibration is stored in the device and displayed the next time the calibration process is activated 6 4 2 Turbidity calib...

Страница 67: ...tings In the connectivity settings see Figure 81 the various types of interfaces on the device can be controlled It is not possible to send several records at the same time through an interface The as...

Страница 68: ...roblems make sure the connectivity settings match the connected devices see Section 6 9 1 Figure 82 RS232 settings 6 5 3 LAN Press the icon to open the overview of the LAN settings Figure 83 If the de...

Страница 69: ...es and spike additions that are stored in the menu are based on German requirements they can always be changed by the user Figure 84 IQC Menue 6 6 1 Monitoring of inspection measuring and test equipme...

Страница 70: ...igure 86 Please note This selection is omitted in the user guide of the NANOCOLOR VIS II The halogen lamp is checked directly Figure 86 Selection of lamps in the lamp check menu After selecting the la...

Страница 71: ...onding value entered via the alphanumeric keypad The data for the input NANOCHECK lot will be saved automatically after the program is run for the first time Alternatively a pre programmed LOT can be...

Страница 72: ...ues are within the specified confidence intervals the test will be given a Passed rating Figure 89 Results of the screening using the NANOCONTROL NANOCHECK Via the icon the result of the test can be p...

Страница 73: ...m with potassium nitrate there is an alternative stray light test available for to the UV range at 220 nm using potassium iodide To perform a stray light test you need the appropriate solutions and a...

Страница 74: ...so www mn net com Press the icon to open the window for specifying the desired standards see Figure 94 Select the entry Standard to open the list of pre defined standards in the device Now either the...

Страница 75: ...ving the cuvette or exiting the measurement menu by means of one of the other icons in the task bar will end the measurement process and save the result in the IQC memory see Section 6 6 7 of the devi...

Страница 76: ...list for standard selection along with the test number and test name see Figure 94 User defined Standard entry Figure 95 Creating a user defined standard Tap the entry Method and then select the meth...

Страница 77: ...d of the multiple determination see Figure 96 Tap the entry Method to open the list of methods available on the instrument Now you can select the desired method from this list The submethods can be se...

Страница 78: ...lculated and maximum diffusion Figure 98 Result of a multiple determination In the example above the average value of the measurements is 117 mg L O2 and the greatest difference of two measurement res...

Страница 79: ...the alphanumeric keypad to enter the maximum permissible deviation in percent in the Margin field This is calculated as part of the measurement based on the absolute measured value difference between...

Страница 80: ...5 Spike additions Through the analysis of spike additions interferences or problems with the sample can be sought and detected Press the icon to open a window for specifyfing the desired standard for...

Страница 81: ...s until the measurement has been carried out Figure 101 Entering the data of a spike addition measurement After inserting and measuring the two sample solutions the result of the measurement will be d...

Страница 82: ...the result in the IQC memory see Section 6 6 7 of the device 6 6 6 IQC counter The IQC counter serves as a reminder to conduct a standard measurement Press the icon to open a window for setting the IQ...

Страница 83: ...type The list box is opened by tapping the entry IQC type in the IQC memory see Figure 104 After selecting an IQC type the IQC memory is selected accordingly Figure 105 Selection of the IQC type Vario...

Страница 84: ...sample depending on the chosen print settings see Section 6 1 6 Figure 106 Display of an IQC single result Details of a measurement can be obtained by selecting a single measured result from the memo...

Страница 85: ...een lines connect measurements within the confidence interval whereas red lines connect out of range measurements When there are more than 25 results the buttons can be used to browse through the indi...

Страница 86: ...e empty Figutre 110 User accounts list Press the icon to create a new user see Figure 111 Enter a user name and password using the text keypad The password has to be entered twice If the administrator...

Страница 87: ...only be deleted by someone with administrator privileges Please note The last user can only be deleted in the Open Mode To change the User Mode press the icon in the user accounts menu and select the...

Страница 88: ...key icon in the status bar If the key is being shown then the user has administrator privileges 6 8 Pictograms To facilitate the use of MACHEREY NAGEL tube tests the pictograms on the lids of test pa...

Страница 89: ...ndow the desired fields can be selected and deselected with the check boxes while the and buttons can be used to change their order On the right hand side of the window are the settings for formatting...

Страница 90: ...121 The data will be transferred as the type configured above 6 9 2 ACRON In order to export to ACRON this menu contains the settings for changing the data export to a format adapted to ACRON For more...

Страница 91: ...t hand corner of the status bar When the timer is activated the time will run down in the status bar and be shown in place of the date and time When the time for the alarm has expired an acoustic sign...

Страница 92: ...sured the latest entered LOT number will automatically be added to the sample information When removing the cuvette the LOT information will be stored together with the result in the result memory see...

Страница 93: ...iles selected or deleted Results can be send to a directly connected printer or can be exported as PNG file to an external storage device and then be printed from another workstation When measurement...

Страница 94: ...the memory menu you can make selections according to various types of sample information as well as according to the measurement method Figure 123 Memory selection according to type of measurement Fig...

Страница 95: ...contains can then be processed further in a spreadsheet program The file name has the following format SNDevice_results csv or SNDevice_result_X csv data of the scan with the number X 7 3 2 Export to...

Страница 96: ...will open Selecting an entry from the list will set the image as the background To add a background press the icon and select an image file from a connected USB storage medium You can delete the list...

Страница 97: ...t Print automatically has been activated in the settings menu for the printer 9 2 Scanner Connect the scanner to one of the instrument s USB A interfaces An acoustic signal will confirm that the scann...

Страница 98: ...uched it might cause severe burnings A halogen lamp provides light for measurements from 320 1100 nm For each standard measurement the lamp emits a corresponding light pulse This results in very low p...

Страница 99: ...ther than the supplied external power supply The warranty is limited to the repair of faulty parts or at our discretion to the delivery of a faultless replacement instrument The recourse to warranty d...

Отзывы: