Cells calibration
The steps for the cells calibration are as following :
1. When the TRITON is fully assembled
2. Make three oxygen flush
3. Close the DSV, open the OPV (Over pressure Valve)
4. Turn on the pPO
2
display
5. Check the altitude
6. Choose the "OXYGEN 98% " calibration (recommended by the manufacturer)
7. Calibrate
8. Check that the voltages are bigger than 40mV then valid the calibration
3.3.5
Fine Calibration
When calibrating oxygen cells with the entire loop, some bad gaz stay inside the Triton.
Calibration is slightly distorted. So you can make a fine calibration, without any tools.
The steps for the cells calibration are as following :
1. Disconnect the inhale counterlung form the scrubber (if needed)
2. Disconnect ADV from exhale counterlung (if needed)
3. Push the ADV into the female connection of the inhale counterlung.
4. Vacuum the circuit with the mouth, and close the DSV.
5. Inject oxygen (no risk of overpressure)
6. Turn on the pPO
2
display
7. Choose the "OXYGEN 99% " calibration (recommended by the manufacturer)
8. Calibrate
9. Check that the voltages are bigger than 40mV then valid the calibration
29