- 121 -
Settings can be done on the following items:
ALM ON/OFF(Alarm switch): Where ―ON‖ is selected, the alarm prompt and saving will
proceed when the IBP (invasive blood pressure) is alarmed; where ―OFF‖ is selected,
alarming will not happen, but
will be prompted beside IBP in the screen parameter
area.
ALM LEV(Alarmlevels): Optional levels are ―High‖, ―Middle‖ and ―Low‖.
ALM REC(Alarm record): select ―On‖ to enable the recorder output when there is any IBP
alarm.
ALARM SETUP: When the alarm measurement value is greater than or less than the alarm
limit , the instrument alarm and prompt .
CH1: ART, PA, CVP, RAP, LAP ,LAP ,ICP ,P1,P2 limit Hi setup:
SYS: Upper SYS alarm limit: used for setting the upper alarm limit
MEA: Upper MEA alarm limit: used for setting the upper alarm limit
DIA: Upper DIA alarm limit: used for setting the upper alarm limit
CH1: ART, PA, CVP, RAP, LAP ,LAP ,ICP ,P1,P2 limit Lo setup:
SYS: Lower SYS alarm limit: used for setting the lower alarm limit
MEA: Lower MEA alarm limit: used for setting the lower alarm limit
DIA: Lower DIA alarm limit: used for setting the lower alarm limit
CH2: ART, PA, CVP, RAP, LAP ,LAP ,ICP ,P1,P2 limit Hi setup:
SYS: Upper SYS alarm limit: used for setting the upper alarm limit
MEA: Upper MEA alarm limit: used for setting the upper alarm limit
DIA: Upper DIA alarm limit: used for setting the upper alarm limit
CH2: ART, PA, CVP, RAP, LAP ,LAP ,ICP ,P1,P2 limit Lo setup:
SYS: Lower SYS alarm limit: used for setting the lower alarm limit
MEA: Lower MEA alarm limit: used for setting the lower alarm limit
DIA: Lower DIA alarm limit: used for setting the lower alarm limit
Caution
Users should guarantee that zero calibration has been done on the sensor before the
measurement; otherwise the instrument has no effective zero value, which may lead to
inaccuracy of measured data.
Once the measured data exceed the alarm limits, the alarm will be triggered.
IBP alarm limits:
Pressure Scale Name
Max Upper Limit
(mmHg)
Min Lower Limit
(mmHg)
Adjustable
Single-Step Length
(
mmHg
)
ART
300
0
1