NIBP specification
Name
Specifications
Measuring mode Automatic oscillation
Measurement
range and
accuracy
Measurement
range for big
Systolic blood pressure
40-270mmHg
Diastolic blood pressure
10-210mmHg
Mean blood pressure
20-230mmHg
Measurement
range for small
Systolic blood pressure
40-200mmHg
diastolic pressure
10-150mmHg
Mean blood pressure
20-165mmHg
Accuracy
5mmHg, when non-invasive BP exceeds the above range,
the monitor can still display normally, but without accuracy
Range
and
accuracy
of
static pressure
Scope:
should be 0mmHg
~
300mmHg
。
accuracy
3mmHg
Range
and
tolerance
of
overvoltage
protection
Big mode
300mmHg
Small mode
150mmHg
tolerance
±3mmHg
Range and error
of alarm preset
Big
Systolic blood
pressure
Upper limit 46-270mmHg
Lower limit 40-264mmHg
Mean
blood
pressure
Upper limit 26-230mmHg
Lower limit 20-224mmHg
diastolic
pressure
Upper limit 16-210mmHg
Lower limit 10-204mmHg
Small
Systolic blood
pressure
Upper limit 46-200mmHg
Lower limit 40-194mmHg
Mean
blood
pressure
Upper limit26-165mmHg
Lower limit 20-159mmHg
diastolic
pressure
Upper limit 16-150mmHg
Lower limit 10-144mmHg
Error
±0.1kPa or±1mmHg (Both maximum)
NIBP
measurement
recall
2000 NIBP measurementdata
(
9
)
TEMP specification
Name
Specifications
Measurement range and
accuracy
Detection range
0
℃~
50
℃
Measurement error
±0.1
℃
Alarm
setting
and
accuracy
Alarm setting range
Upper limit of alarm should be 0.1
℃~
50.0
℃
Lower limit of alarm should be 0
℃~
49.9
℃
Alarm error
±0.1
℃
Display Resolution
0.1
C