5
Figure 1-2 Main screen display
Information Area (
①
):
Prompt message in the information area appears and disappears together with the reported state. In
accordance with the content, it is divided into:
1.
Network setup: Cursor stops in
sign, press to enter system setup.
Network Bed No.: monitor with central monitoring system connected network bed No.
IP Address: IP is 200.200.200.X,
(
X is Network Bed No.:, from 1 to 100
)
Attention
Connection state of CMS (central monitoring system) : when displaying
, no connection; when
displaying
, connection.
Network Bed No. should not conflict with other monitors, or there will be unnecessary problems for
the monitor.
If any problem happens because of confliction, unplug network wire, restart the monitor, reset the
Network Bed No. and connect the network wire again.
2.
Patient Information: press to enter the Patient Information menu when cursor is in bed No., name or
patient group area.
Bed No.: Refer to the bed number of the patient being monitored, this area will be blank if there
is no input.
Name: Patient name, this area will be blank if there is no input.
Sex: Male and Female.
Weight(kg): patient‘s weight
Patient Type: big or small
Information Area
Parameter
Area
Waveform
Area
Menu Area