70
PVS bigeminy No pacing PVS bigeminy
Selectabl
e by users
PVS trigeminy No pacing PVS trigeminy
Selectabl
e by users
R ON T
No pacing
HR<100, RR interval < 1/3of average interval, after which
there is a compensation interval of which length is 1.25 x R-R
mean (next R waveform is moved up to the previous T wave.)
Selectabl
e by users
Single PVS
No pacing Single PVS not belonging to the above types
Selectabl
e by users
Tachycardia
Full
Successively 5 QRS waveforms and RR mean smaller than
0.5s
Selectabl
e by users
Bradycardia
Full
Successively 5 QRS waveforms and RR mean greater than 1.5s
Selectabl
e by users
MB(Missed
Beat)
No pacing
Where HR is smaller than 100times/minute, HB is not detected
in the 1.75 times of time for RR mean; or where HR is greater
than 100times/minute, pulsation is not detected within 1
second.
Selectabl
e by users
Pacemaker
doesn‘t work.
Pacing
In the 1.75 times of time for RR mean there are no QRS
waveform and pacing pulse (only for the patient with the
pacemaker).
Selectabl
e by users
Pacemaker
doesn‘t capture
pacing.
Pacing
In case of pacing pulse there is no QRS wave in the 1.75 times
of time for RR mean (only for the patient with the pacemaker).
Selectabl
e by users
Applied Patient Type: ―Full‖ means the arrhythmia analysis that can be given to the patient with the
pacemaker as well as that to the patient without the pacemaker.
―Non-pacing‖: means the arrhythmia analysis that can be given to the patient without the pacemaker.
―Pacing‖: means the arrhythmia analysis that can be given to the patient with the pacemaker.
Prompt Message:
Prompt
Message
Cause
Alarm
Level
ARR Learning
QRS wave template necessary in ARR analysis is forming.
No Alarm
Caution
Arrhythmia name displays in the alarm message area.
11.6
10.6 ECG Alarms & Reminders
Alarming information
Possibly happened alarms during ECG measurement are classified into physical alarms and technical
alarms, while during the same process there may be various reminders. The visual and audio appearance of
the monitor upon happening of such alarms or reminders, are as described in the Alarming chapter. On the
display screen, physical alarms and general reminders are displayed in the alarming area, while technical
alarms and those reminders unable to activate alarms are displayed in the information area. In this chapter
all the descriptions on alarming applies out of the arrhythmia and ST segment analysis parts.
When the alarming record function under relevant menu is turned on, those physical alarms activated