60
Figure 11-5 ECG
Setup
Alarm ON/OFF: Select ―ON‖ to give alarm prompt and storage when the heart rate alarm
happens.
will be prompted beside ECG.
Alarm record: Users can select ―On‖ to print HR alarms when they happen
ST ALM ON/OFF: Select "ON‖ in the event of ST1 or ST1 overrun alarm prompt and storage,
select the "OFFs not an alarm, and ST1 next have "
‖ pompt.
PVCs ALM ON/OFF: Select "ON‖ in the event of PVCs or PVCs overrun alarm prompt and
storage, select the "OFFs not an alarm, and PVCs next have "
‖ pompt.
HR ALM SETUP: setup heart rate hight limit ,middle limit ,low limit of the upper and lower
limits .
Alarms will happen once the HR values exceed the upper or lower limit.
Adjustable ranges for HR alarm upper & lower limits are as follows:
Name
Highest upper limit
Lowest lower limit
Adjustment
step
length
Big /small
350
15
1
Attention
Users should set the alarm upper & lower limits based on the clinical conditions of every patient.
Setting of the HR alarm upper limit is very important, and users should not set it too high but
consider fluctuation factors. The set HR alarm upper limit should not be over 20 beats/ minutes
than patient HR.
11.5.2
ECG setting in waveform area