![LUMENIS PULSE 100H Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/lumenis/pulse-100h/pulse-100h_operators-manual_1948102074.webp)
Lumenis
®
Pulse
TM
50H / 100H Laser Systems
Electromagnetic Immunity
UM-20006520DE, Rev. A
Page 74
Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Lumenis Pulse 50H & 100H systems are intended for use in the electromagnetic environment specified
below. The customer or the user of the Lumenis Pulse 50H & 100H systems should assure that it is used in
such an environment.
IMMUNITY test IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 2 kV
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 1 kV differential
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
± 2 kV common mode ± 2 kV common
mode
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% UT
(>95% dip in UT) for
0,5 cycle
40% UT (60% dip in
UT) for 5 cycles
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 s
N/A
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Lumenis Pulse 50H or 100H systems
requires continued operation during
power mains interruptions, it is
recommended that the Lumenis
Pulse 50H or 100H systems be
powered from an uninterruptible
power supply.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.