Manual Reset
If laser emission is interrupted during treatment (e.g., main electrical
power loss), the laser system automatically turns
Off
. To resume
treatment, you must manually restart the laser system using the main
On
/
Off
button.
Electronic Fault Detection Circuitry
If any of the electronic system monitors detect a fault condition, laser
emission is disabled. The high voltage power supply disables, the safety
shutter closes, and the footswitch disables.
Safety Interlocks
The laser system has a safety interlock on the optical fiber laser connector.
Precision of Displayed Values
The precision of the energy and rate values displayed on the control
screen are factory preset to within ±5% of a calibrated standard. The
energy of every pulse is monitored by two internal detectors to ensure that
no safety hazard is caused by failure of a single component. If the
delivered system energy deviates from the commanded parameters by
more than 20%, you are notified and can continue lasing following
acknowledgment. Following 5 such occurrences in a single session this
becomes a fatal error and lasing cannot continue (laser system shuts
down).
Space Requirements
Position the laser console a minimum of 50 centimeters (20 inches) from
walls, furniture, or other equipment.