Lumenis
®
Pulse
TM
50H / 100H Laser Systems System
Overheats
UM-20006520DE, Rev. A
Page 54
System Overheats
The laser system may overheat if it is used at a high power for an
extended amount of time.
• Verify that the treatment room temperature is between 10 and 30°C
(50 and 86°F).
• Verify that the laser system is at least 50 centimeters (20 inches)
from walls, furniture, or other equipment.
NOTE:
If the laser system overheats, do not turn it off. Leaving the laser system
on allows the internal cooling system to quickly cool the system’s internal
components. Allow the system to cool for several minutes and resume
normal operation.
Message Appears: Attach an Authorized Fiber
SIS enabled Lumenis Pulse 50H and Lumenis Pulse 100H laser systems
will only operate with Lumenis-qualified SIS (Secure Identification
System) optical fibers. Attaching any other type of fiber will disable laser
emission.
• Attach an authorized fiber or Contact Lumenis Customer Service to
obtain the correct fibers.
• Refer to the optical fiber’s Directions for Use document for specific
instructions on how to work with the fiber.
Message Appears: Attach fiber
The optical fiber’s connector is not properly connected to the laser
system.
• Connect the optical fiber as instructed earlier in this manual.
Message Appears: Attach footswitch
The footswitch is not properly connected to the laser system.
• Connect the footswitch as instructed earlier in this manual.